9.6
Maintenance
Menu "Maintenance" serves to retrieve the following submenu items:
■
Reset System: Restart system
■
Maint Mode: Set Maintenance mode
■
Test Analog Out: Check the power value on the analog output
■
Test Relay: Relay test
■
Reset Parameter: Reset parameters to the default setting
Measuring
T=150 °C
Measuring T=150°C
0
1000
CO
564
ppm
Parameters
back
Reset System
Maint Mode: No
Maintenance
Test Analog Out
Test Relay
maint
Reset Parameter
Figure 59: Path from Measuring Screen to change parameters (example)
9.6.1
Measuring system restart
Submenu item "Reset System" serves to restart the system.
b
Switch to the Service menu with "maint",
.
NOTE
•
A restart cannot be interrupted or terminated.
•
The restart takes a few seconds.
•
An input is not possible.
Parameters
back
Reset System
Maint Mode: No
Maintenance
Test Analog Out
Test Relay
Parameters
back
Reset System
Start: Enter
Are you sure to
start reset
procedure ?
Parameters
Reset System
Reset System
Please wait !
************
Measuring
T=150 °C
Measuring T=150°C
0
1000
CO
564
ppm
Figure 60: Restarting the system
9.6.2
Activating/deactivating maintenance mode
When maintenance mode is active, the output relay drops off and the analog output
holds the last value.
NOTE
Select "Yes" and press "Enter" to activate maintenance mode.
The "Service" LED is on as long as maintenance mode is activated.
MENUS
9
8008932/16YT/V4-0/2022-02 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM901-02
73
Subject to change without notice