9.4.2
Setting the humidity correction
Submenu item "Normalization" serves to set whether the measured value is output
calculated as "wet" or "dry". The default value entered for H
2
O is used here,
the humidity content in sample gas", page 63
.
The damp correction is activated when "wet" is set. Available are:
•
dry
•
wet
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
Parameters
back
Mode
:
dry
Normalization
edit: Enter
select
Parameters
back
Mode
:
Normalization
dry
wet
select
Parameters
back
Mode
:
Normalization
dry
wet
Parameters
back
Mode
:
wet
Normalization
edit: Enter
Figure 33: Selecting the humidity correction
9.4.3
Changing the response time
Submenu item "Response Time" serves to change the response time.
NOTE
Basic factory setting: 24 s.
Allowable values: 5 ... 360 s.
The actual total response time is 30s because the preset value (11) of the Median filter
extends the response time by 6 s.
Contact SICK Service when you are not sure which response time should be set.
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
Parameters
back
Response Time
Time: 25 s
select
Parameters
back
Time: 25 s
Response Time
edit:Enter
Parameters
back
Response Time
Time: 25 s
select
Parameters
back
Response Time
Time: 24 s
select
Parameters
back
Time: 24 s
Response Time
edit:Enter
Figure 34: Changing the response time (example)
MENUS
9
8008932/16YT/V4-0/2022-02 | SICK
O P E R A T I N G I N S T R U C T I O N S | GM901-02
57