4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
back
Scale Low :
0 °C
Scale High :
250 °C
Edit: Enter
Temperature Input
back
Temperature Input
back
Temperature Input
Select
Select
Input low :
4,0 mA
Input High :
20,0 mA
Scale Low :
0 °C
Scale High :
250 °C
Input low :
4,0 mA
Input High :
20,0 mA
Scale Low :
0 °C
Scale High :
250 °C
Input low :
4,0 mA
Input High :
10,0 mA
back
Temperature Input
Scale Low :
0 °C
Scale High :
250 °C
Input low :
4,0 mA
Input High :
10,0 mA
Edit: Enter
Figure 43: Changing the value for the upper output value (example)
9.4.8
Setting the humidity content in sample gas
Submenu item "Humidity" serves to set the value for the humidity content in the sample
gas as default value.
NOTE
Factory basic setting: 0.0 %.
This value is used to perform a gas dryness correction.
Max. possible value: 99 %.
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
back
Substitute: 0,0 %
Edit: Enter
Humidity Input
back
Humidity Input
back
Humidity Input
Select
Select
back
Humidity Input
Edit: Enter
Substitute: 00,0 %
Substitute: 30,0 %
Substitute: 30,0 %
Figure 44: Entering the default value for humidity content in sample gas (example)
9.4.9
Setting the sample gas pressure
Submenu item "Pressure" serves to set the default value for the sample gas pressure.
The measuring system uses this value for correction.
The pressure correction is made in ppm or Norm in the display.
NOTE
Factory setting: 1013 hPa
Min. value: 800 hPa
Max. value: 1200
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".