7
Temperature:
see "Parameter settings for the external temperature sensor
.
8
Humidity:
see "Setting the humidity content in sample gas", page 63
.
9
Pressure:
see "Setting the sample gas pressure", page 63
.
10 Analog Out:
see "Changing the Live Zero value of the analog output (Analog Out)",
.
11 Calibration:
see "Changing the calibration values", page 64
.
Measuring
T=150 °C
Measuring T=150°C
0
1000
CO
564
ppm
Parameters
back
Physical
Normalization
Parameter Settings
Response Time
MeasuringRange
Parameters
Settings
Device
Parameter
Service
Unit
Parameters
Unit : mg/ Nm³
Physical Unit
edit:Enter
back
back
Figure 31: Path from Measuring Screen to change parameters (example)
Changing parameter settings
b
Press "Enter" in the display for current parameter setting.
b
Enter the password,
8.3 "Entering the password", page 52
b
Press "Left arrow" after entering the new value.
Menu "Parameter Settings" is displayed again.
9.4.1
Changing the physical unit
Submenu item "Physical Unit" serves to set the physical unit for the CO value output.
Available are:
•
ppm
•
mg/m
3
N
•
mg/m
3
1
Follow the menu path as described,
2
Enter the password,
see "Entering the password", page 52
3
Enter the changes.
4
Confirm value with "Enter".
5
Press the "Left arrow" to return to menu item "Parameter Settings".
Parameters
back
Unit : ppm
Physical Unit
edit:Enter
Parameters
back
Physical Unit
select
ppm
mg/Nm³ mg/m³
Parameters
back
Physical Unit
edit: Enter
Unit : ppm
Unit : mg/ Nm³
Parameters
back
Physical Unit
select
ppm mg/Nm³ mg/m³
Unit : mg/ Nm³
Figure 32: Changing the physical unit
9
MENUS
56
O P E R A T I N G I N S T R U C T I O N S | GM901-02
8008932/16YT/V4-0/2022-02 | SICK