Terminal Modules (Accessory)
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
69
7.1.2
Installation
1
Secure the terminal module to a mounting rail.
2
Connect the plug connector X1 of the terminal module with the “Analog and digital I/O”
plug connector of the operating unit. To this purpose use the special connection cable
(order no.
Connect the “Connecting module” plug connector to the terminal module.
Connect the “Remote unit” plug connector to the operating unit.
3
Check the possible configurations of the terminal module (
4
Establish the signal connections as required (
).
7.1.3
Configuration options (jumper)
Use of terminal strip X10 or X20
Several signal connections can be used either on terminal strip X20 or X10. Two plug-in
jumpers (jumper
) serve for configuration purposes.
For connection on X20:
Close the respective plug-in jumpers (position jumpers).
For connection on X10:
Open the respective plug-in jumpers (remove jumpers).
Feedback loop for “Malfunction” and “Alarm”
A feedback loop can be set up for the “Malfunction” and “Alarm” signal outputs.
Application example:
A relay is controlled with the signal of the operating unit; the
potential-free relay switching contacts are provided on the feedback signal connection.
To use the “Malfunction” and “Alarm” signal outputs directly:
Close the respective plug-
in jumpers (position jumpers
To use the feedback loop:
Open the respective jumpers (remove the jumpers).
The connector pin assignment of the connection cable is not identical.
Do not confuse the ends of the connection cable.
Содержание EuroFID
Страница 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Страница 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Страница 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Страница 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Страница 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Страница 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Страница 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Страница 155: ...Shutdown EuroFID Operating Instructions 8011094 V2 3 SICK AG 155 EuroFID 13 Shutdown Preparations Shutdown procedure ...
Страница 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Страница 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...