5.1
General
The maintenance work to be carried out consists of:
●
Cleaning work (
●
Securing the purge air supply function (
●
Checking/correcting the alignment of the optical axes of sender/receiver unit and
reflector (
).
Take the following steps to set the measuring system to "Maintenance" mode before
starting maintenance work.
Connect the measuring system to the laptop/PC using the USB cable and start
SOPAS ET.
Click button "Network Scan" in tab "Network Scan Assistant", select device file "MCU"
and move it to window "Project Tree" (
Switch to the "Maintenance / Maintenance" directory, activate the "Maintenance on/
off" checkbox in the "Maintenance / Operation" group and click "Set State" (see
Figure 84
Setting "Maintenance" mode
Resume measuring operation after completing the work (deactivate the "Maintenance on/
off" checkbox in the "Maintenance / Operation" window and click "Set State").