115
8024493/AE00/V1-0/2020-09| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100 Ex-2K
Subject to change without notice
TROUBLESHOOTING
13
13.6.3
Replacing the fuse
1
Disconnect the control unit from the power supply and observe the waiting time before
opening the door.
2
Take suitable precautions to prevent dust from entering the enclosure.
3
Open the control unit door.Remove fuse holder (1) and open.
4
Remove the transparent fuse cover.
5
Replace defective fuse (2) (
see “Other accessories”, page 127
6
Close and attach the fuse holder.
7
Close the door. Switch the power supply on again.
Fig. 56: Exchanging the fuse (shown with fuse cover removed)
13.6.4
Sending in devices
Information on sending a device to the factory for inspection or repair,
WARNING:
Risk of explosion when using an unspecified fuse
There is a risk of explosion when a different type of fuse is used.
▸
Only use fuses of type 522.723 from Eska.
Fuse holder
Fuse