START-UP AND PARAMETER SETTINGS
4
4.2.2
Changing settings using the jumper on the processor board
The following steps are then necessary (
see “Adapting the sender/receiver unit to the duct
):
▸
Loosen CAMLOCK clips of the sender/receiver unit, swivel the electronics unit to the
side and take off from the mounting flange.
▸
Loosen the fastening screws on the electronics unit, carefully pull the electronics unit
out of the cover.
▸
After changing the settings, reassemble the sender/receiver unit in reverse sequence.
4.2.2.1
Setting measuring range and limit value
▸
Plug the jumper in the slot for the setting desired for the measuring range and limit
value.
Fig. 22: Processor board setting options
Parameter
Slot 1
Slot 2
Slot 3 (default)
Value at LZ
0 SI
0 SI
0 SI
Value at MBE
100 SI
500 SI
3000 SI
Limit value
50 SI
250 SI
1500 SI
Slot 4 is not used.
Microswitch for baud
rate Modbus
®
Rotary switch for
setting measuring
range
Microswitch for
addressing