13
© Copyright Siborg Systems Inc.
used for impedance values lower than 1 kOhm. For inductance
only serial mode is used unless parallel mode is explicitly
specified.
The device will beep when the resistance is less than 10
Ω
, or a
short circuit is found.
This setting must be selected in
the menu when evaluating
electrolytic capacitors for
rejection by ESR value. The
voltage value should be taken from the capacitor’s housing.
If
OFF is selected, the capacitor parameters such as the capacitor
quality according to the table and nominal voltage will not be
shown on the display
When selected, the device will hold measurement data on the
screen. A beep will sound when the measurement is made, and
the values are held on the screen after the component has been
disconnected from the device. The device will only reset when a
new component is measured, or the test probes are shorted.
In order to set a default measurement mode that will be
activated when the device is turned on next time, select the
mode, press the joystick and hold for 2 beeps. This measurement
set
ting will be stored in the device’s non
-volatile memory for all
other measurements until selecting other parameters or Default
in the menu.
By selecting Default, the R-L-C-D mode parameters are reset to:
•
Primary Parameter is Auto
•
Secondary Parameter is Auto
•
Frequency is Auto
•
Signal Level - 1.0 Vrms
•
Measurement Time: 0.5 seconds
•
Equivalent Scheme is Auto
•
Sound is On
•
C-voltage is Off
•
Data Hold is Off
If you hold the joystick up to 2 beeps, then all parameters are
stored in the non-volatile memory and will be loaded when the
device is powered on.