background image

 

Measurement Procedure

 

(1) Measure and record the mass of the PTFE binder filter using a balance. 
(2) Immerse the metal collection plate in a neutral cleaning fluid, remove dirt and contamination from 

the surface, wash it with water, rinse it with purified water, then immerse it in alcohol and wipe it 
with degreased gauze. 

(3) Take the grease with a micro spatula as shown in Fig. 2. At this time, sufficiently stir the grease in 

the container before use. 

 
 
 
 

 

 

Fig. 2

 

Fig. 3

 

Fig. 4

 

The grease recommended by Sibata (080150-011 and 080150-012) is a mixture of two types.   

They can separate during storage, so mix well before use. 

(4) Place the grease guide on top of the metal collection plate as shown in Fig. 3, and place the 

grease substantially in the center. 

(5) Level the surface of the grease using a spreading spatula as shown in Fig. 4. 
(6) Gently remove the grease guide from the metal collection plate, then accurately weigh the metal 

collection plate. (When not evaluating that the 4 

μ

m dust has been removed by 50 % or more, it is 

not necessary to weigh the metal collection plate.) 

(7) Set the weighed PTFE binder filter and the metal collection plate in the holder in the order shown 

in Fig. 1. 

(8) With the holder placed in the suction pump, adjust the aspiration rate to 2.5 L/min and collect dust. 
(9) When dust collection is completed, carefully bring back the holder, and weigh the PTFE binder 

filter. (When evaluating that the 4 

μ

m dust has been removed by 50 % or more, also weigh the 

metal collection plate.) 

 

Grease (sold separately) 

In order to prevent re-entrainment of particles from the surface of the metal collection plate on 
which particles PM4 (removing 50 % of 4 

μ

m particles) or more were collected, apply grease. The 

air that passes the particle size separator has extremely high velocity, and the grease must not be 
allowed to flow and disperse as a result of this high flow velocity. Therefore use a high viscosity 
grease. Sibata has prepared the recommended grease together with equipment such as a 
spreading spatula for application, etc. 

 

 

Items Sold Separately

 

 
 
 
 
 

20.06.11K (10) 

Item Code 

Product Name 

Qty 

080150-011 

Sampler Grease Set (silicone grease, micro spatula, and spreading spatula) 

080150-012 Silicone 

Grease 

080130-098025 

Glass Fiber Filters, PTFE Binding, TF98R, 25 mm dia. (

PTFE binder filter) 

100 

080860-304 

Mini Pump  MP-

300N

II 

080860-5050 

Mini Pump  MP-W5P 

Grease guide 

Place on top 

Metal

 

collection plate 

Grease

 

Grease

 

Отзывы: