- 26 -
Data Logger Setting Screen (LOG)
This screen is used to specify settings for logging measurement data in the product’s internal
memory.
Record ON/OFF:
Each touch switches log recording [ON]/[OFF] and determines whether
the data logger is enabled or disabled.
Record Data
One Time/Overwrite: Each touch switches between the [OT] (one-time) and [OW] (overwrite)
modes. The following describes the process assuming no remaining
memory is available.
OT: Recording stops when no memory is available.
OW: Past data is overwritten when no memory is available. (Past data is deleted.)
The [MEMORY] value displayed indicates the number of data points currently recorded in the
fiber monitor. A maximum of 5450 data points can be recorded.
Touching the [DEL] button deletes all recorded data. Once the data is deleted, it cannot be
recovered.
When finished configuring settings, touch the [ENTER] button to return to displaying the sub-
menu screen again.
In the unlikely event some sort of problem occurs, note
Sibata accepts no responsibility for providing compensation
for data that could not be acquired or recorded, loss of data
or other content, or for any other direct or indirect damages
involved. Be sure to back up data regularly to be prepared in
case of an accident or failure. Sibata recommends verifying
functionality in advance and periodically inspecting the
product.
Data Recorded in Internal Memory
• Equipment serial No. (S/N)
• Log date
• Log time
• Fibrous particle converted concentration value (f/L)
• Accumulated count value (count)
• Accumulated flowrate (L)
• Elapsed measurement time (min)
• Respective alarms (AL1, AL2, AL3, AL4, AL5, AL6, and AL7)
Each line of recorded data uses one data point. A total of 5450 lines of data can be recorded.