OPTION INTERFACES
135
OPTI
ON INTERFACES
Chapter
7
7-1.
BCO Option
You can set data to be output as fixed, or read data by setting external selection.
■
BCD output data list
*1
BCD data select input condition when the BCD Output Data Selection is External.
*2
Value selected by Comparison Standard in Comparison Setting of the measurement work
no. specified by external input.
BCD data select 1
*1
BCD data select 0
*1
Standard
Before Value
Double Hold
OFF
OFF
Real Time value Real Time value Real Time value
OFF
ON
Hold value
Hold value
Hold value
ON
OFF
Real Time value Difference Value
Hold value
ON
ON
Real Time value Real Time value Real Time value
BCD data select 1
*1
BCD data select 0
*1
Double Hold &
Before Value
Comparison
Double Hold &
Relative Value
Comparison
Double Hold &
Before Value
Comparison &
Relative Value
Comparison
OFF
OFF
Real Time value Real Time value
Real Time value
OFF
ON
Hold value A
Hold value (A/B)
*2
Hold value (A/B)
*2
ON
OFF
Hold value B
Relative value
Relative value
ON
ON
Difference Value Real Time value
Difference Value
7
OPTION INTERFACES
A1
-
COM
COM
-
B1
A2
OUT
1
1000
OUT
B2
A3
OUT
2
2000
OUT
B3
A4
OUT
4
4000
OUT
B4
A5
OUT
8
8000
OUT
B5
A6
OUT
10
10000
OUT
B6
A7
OUT
20
20000
OUT
B7
A8
OUT
40
40000
OUT
B8
A9
OUT
80
80000
OUT
B9
A10
OUT
100
Minus (Polarity)
OUT
B10
A11
OUT
200
OVER
OUT
B11
A12
OUT
400
STAB
OUT
B12
A13
OUT
800
STROBE
OUT
B13
A14
IN
BCD Data Hold
Logic Select
IN
B14
A15
IN
BCD Data Select 0
BCD Data Select 1
IN
B15
A16
-
-
B16
- Connector pin assignments for BCD output
FCN-365P032-AU/FCN-360C032-B