CABLE CALIBRATION
If CALIBRATE has been chosen the unit will display if a cable is connected
CHECK CABLE REEL
IS 305 METERS
PRESS SET
TO CONTINUE
Or if no cable is connected:
No cable
CONNECT 305
METER REEL
Once a 305 meters cable reel is connected press SET, the display changes to
CABLE RTN TIME
2444.9 ns
*
PRESS SET
TO CONTINUE
Pressing SET will calculate the cable calibration factor and display.
CAL FACTOR
8.016
*
PRESS SET
TO CONTINUE
* Note: numbers taken from Digitek Quad Shield RG6
The calibration factor will automatically be set for this unit and if required this value can be loaded into
ShotTrack for Windows for future use with this cable. Generally the cable produced now is of so good quality
that this is only needed to be done once for each new cable type used.
If a 305 Meter reel is not available
then the same procedure should be used until the cable return time is
displayed. This value should be written down and 25 meters of cable cut from the reel. The new return time
written down and the OFF Button pressed to get out of the function.
Then this formula should be used to calculate a calibration factor which must be entered into the unit using
ShotTrack for Windows.
Calibration factor = First reading – Second reading / 25