Shortridge Instruments Airdata ADM-870 Скачать руководство пользователя страница 38

ADM-870 05/23/08

34

11.0  FLOWHOOD OPERATING PROCEDURE

The meter handle should be removed from the AirData Multimeter by unscrewing the captive fasteners on the

sides.  The meter is inserted into the recess in the FlowHood base, using the captive fastener inside the base

to secure it.  The two pneumatic tubes from the flow sensing grid attach to the pneumatic inlets on the meter.

The tube from the top side of the grid connects to the positive (+) pneumatic inlet on the meter; it is very

important that these connections are not reversed, since this will cause invalid flow measurements.  The flaps

plug inserts into the flaps jack on the back of the meter.

The thumbswitch plug connected to the handgrip is inserted into the external read jack on the left side of the

meter.  The TemProbe sensor is plugged into the temperature input jack on the back of the meter through the

cutout designed into the FlowHood base.

CAUTION

Avoid possible damage by removing all three electrical connections and both pneumatic connections before

removing the meter from the FlowHood base.

After being turned on, the meter will flash a full set of eights to test the display, and then pass through a brief

CAL phase.  The display will then read FLO-HOOD, followed by CF: READ.  After the operator presses the

READ key, either CF: UNCORR or CF: CORR will be displayed during the measurement period.  CALC will

be displayed during the brief calculation period, followed by the results.  Each subsequent measurement is

triggered by pressing the READ key.  Pressing the READ key also clears all previous data from the registers

and the display.  The meter will automatically turn off if not used for eight consecutive minutes unless readings

are stored in memory.

Place the FlowHood over the outlet (or inlet) diffuser or grille, so as to capture and direct the air flow through

the FlowHood base.  The foam gasket around the edge of the top assembly should be firmly pressed against

the ceiling or the edges of the diffuser.  Air flow through the unit should not be blocked or influenced by any

object within one foot of the FlowHood base.  The unit should be supported with both hands, one on the handle,

and the other at any convenient point on the outside of the base.  Set the flaps control knob to the open

position.

Press the READ key or the thumbswitch on the FlowHood handle and the reading will be displayed directly in

cfm or litres/sec.  A negative number indicates that the air flow is being exhausted through a return outlet (or

that the hoses are crossed).  A second reading with the flaps closed will be necessary to determine the

backpressure compensated air flow.

SAFETY NOTICE

When using the FlowHood to test air flow at ceiling outlets, be sure that you can safely raise and support the

unit throughout the measurement.   This is especially important when working on a ladder.

11.1  AIR FLOW - NONBACKPRESSURE COMPENSATED READINGS

See Section 9.2 BACKPRESSURE COMPENSATION for a general discussion of backpressure and its effect

on any capture hood device before using the FlowHood for nonbackpressure compensated readings.

This technique measures air flow without compensating for FlowHood induced backpressure.  This method is

used for both supply and return outlets.

After the meter and all of its connections have been attached to the FlowHood, as described previously, set

the flaps control knob to the open position.  Place the FlowHood over the outlet (or inlet) diffuser or grille.  Wait

two or three seconds to allow the air flow to stabilize, and then press the READ key or the thumbswitch on the

FlowHood handle.  The displayed result will represent the flow value of the outlet while the FlowHood is in

place.  This flow reading will generally be somewhat less than the actual flow, due to the backpressure effect

of the FlowHood.

Readings as low as 23 cfm may be taken with the flaps open.  Readings as low as 17 cfm may be taken with

the flaps closed (see section 12.5 LINEAR SLOT DIFFUSERS).  The meter will not permit nonbackpressure

compensated readings of more than 500 cfm if the flaps are closed.  The flaps must be opened to continue.

11.2  AIR FLOW - BACKPRESSURE COMPENSATED READINGS

Air flow through an outlet (or inlet) is reduced to some degree by any capture hood device placed over the

opening.  Correction curves had traditionally been used to provide an average correction factor for the

backpressure effects at different air flows.  This method measures air flow with compensation for the FlowHood

induced backpressure automatically calculated into the displayed result.

Содержание Airdata ADM-870

Страница 1: ...nts Inc 7855 East Redfield Road Scottsdale Arizona 85260 3430 Phone 480 991 6744 Fax 480 443 1267 www shortridge com S i AIRDATA MULTIMETER ADM 870 ELECTRONIC MICROMANOMETER AIR FLOW VELOCITY PRESSURE...

Страница 2: ...IL 18 6 3 3 EXHAUST HOODS AIRFOIL PROBE 19 6 3 4 LAMINAR FLOW WORKSTATIONS AIRFOIL PROBE 19 6 4 VELGRID AIR VELOCITY 19 6 4 1 CHEMICAL EXHAUST HOODS VELGRID 20 6 4 2 LAMINAR FLOW WORKSTATION VELGRID 2...

Страница 3: ...L PRESSURE FUNCTION 40 14 3 ABSOLUTE PRESSURE FUNCTION 40 14 4 AIR FLOW ACCURACY 40 14 5 DUCT TRAVERSE COMPARISONS INCLINED MANOMETER MICROMANOMETER 40 14 6 DUCT TRAVERSE USING THE AIRDATA MULTIMETER...

Страница 4: ...6 4 VELGRID ASSEMBLY 21 7 1 STATIC PRESSURE PROBE 23 8 1 ADT442 TEMPROBE 25 8 2 AIRDATA MULTITEMP 26 10 1 FRAME STORAGE 30 10 2 FLOWHOOD IN CASE 30 10 3 FLOWHOOD ASSEMBLY 31 10 4 2X2 FRAME ASSEMBLY 32...

Страница 5: ...emperatures can be measured from 67 F to 250 F either individually or in conjunction with an air flow or velocity measurement The Model ADM 870 AirData Multimeter performs these additional functions w...

Страница 6: ...readings AIR DENSITY CORRECTION The air density correction range is 14 40 in Hg and 32 F to 158 F for correction of air flow and velocity measurements The readings represent either local density air...

Страница 7: ...e accurate measurements Bleedthrough typ 0 0004 cu in in wc measurement Quiescent bleedthrough max 0 0005 cu in in wc minute TUBING The maximum recommended length of pneumatic tubing for the measureme...

Страница 8: ...emory mode then alternate action for average and for total This key was labeled MEM prior to May 1993 ON Turns the meter on RCL Recalls stored readings in reverse order RCL Recalls stored readings in...

Страница 9: ...e meter is dropped and may cause the keypad to become nonfunctional until the meter is reset If the meter continues to fall into lockout it may have been damaged and should be returned for repair PNEU...

Страница 10: ...EAD Read Hold for Speed or Halt RCL Recall in Entry Order RCL Recall in Reverse Order OFF Power Off ON Power On Positive Pressure Port Negative Pressure Port Flaps Jack for FlowHood Flaps Plug Threade...

Страница 11: ...hat the temperature readings will be displayed in degrees Celsius C LS READ This display indicates that the meter has been placed in the air flow function and will read in litres per second L s S READ...

Страница 12: ...mode is selected and also upon each subsequent operation of the READ key AUTOSTOR This message indicates that the meter has been placed in the automatic memory mode which integrates the automatic rea...

Страница 13: ...de is selected or when the meter is first turned on if the flaps plug is connected HALT This message will be displayed when an automatic reading sequence has been halted manually by holding down the R...

Страница 14: ...tions do not apply to SPEED readings STD 70 F This message will be flashed during air flow or velocity measurements performed without the TemProbe The resulting flow or velocity value will be calculat...

Страница 15: ...ADM 870 05 23 08 11 VELGRID This signal will be flashed when the VelGrid mode is selected and also upon each subsequent operation of the READ key...

Страница 16: ...ited by pressing the SHIFT key followed by the CLEAR key 5 2 MEMORY AVERAGE TOTAL FUNCTION The memory average total function may be used with any of the measurement modes and allows the storage of up...

Страница 17: ...ay be combined to gain the increased resolution afforded by the increased time base of the measurement interval This function permits the operator to store up to 100 fully automatic repeated readings...

Страница 18: ...he SPEED READ function may be terminated while HALT is displayed by pressing the SHIFT key followed by the CLEAR key or by turning the meter off The memory function may be activated while HALT is bein...

Страница 19: ...ating air flow from velocity measurements at most air movement devices such as Clean Room HEPA filters chemical exhaust hoods safety cabinets laminar flow work stations coil and filter face velocities...

Страница 20: ...90 degree angle located near the base The measurement range of the AirData Multimeter with the pitot tube is 25 to 29 000 fpm calibration accuracy is certified to 8 000 fpm The stainless steel pitot t...

Страница 21: ...city pressure rather than velocity A pitot tube used with the AirData Multimeter reads out directly in velocity when used in the PITOTUBE mode and reads velocity pressure when used in the DIF PRES mod...

Страница 22: ...tally and up against the bottom edge of the sash door Tape markers on the AirFoil probe and along the sash door edge at 6 centers will aid in accurate positioning of the AirFoil probe The individual r...

Страница 23: ...nt is usually adequate 6 4 VELGRID AIR VELOCITY IMPORTANT See Section 6 1 VELOCITY CORRECTION FACTORS The VelGrid accessory is designed especially for use in the measurement of general face velocity c...

Страница 24: ...panel face The velocity average obtained in this manner can be used to calculate the volumetric air flow rate as described in Section 6 4 3 AIR FLOW CALCULATION 6 4 3 AIR FLOW CALCULATION FROM VELGRID...

Страница 25: ...ADM 870 05 23 08 21 6 4 VELGRID ASSEMBLY...

Страница 26: ...ity conditions other than standard must be corrected for local air density conditions if these results are to represent true velocity Air velocity readings taken with the ADM 870 in the standard densi...

Страница 27: ...RESSURE PROBES Static pressures are measured with static pressure probes These probes are brass colored and have a single tubing connection in the magnetic base Connect the tubing from the static pres...

Страница 28: ...the atmosphere and the meter is in the absolute pressure mode The measurement range is 10 40 in Hg Specific absolute pressure source measurements may be obtained by connecting the pressure source dir...

Страница 29: ...emp combines with the memory function of the AirData Multimeter to store up to 100 temperature readings along with the sequence tag for each reading Each temperature reading may be entered into memory...

Страница 30: ...lugs or direct immersion Some applications include humidity control systems direct expansion A C systems outside air adjustment hot water heating chilled water condenser water and many other circulati...

Страница 31: ...n the display immediately prior to the display of the air flow reading The flow is calculated using an assumed standard temperature of 70 F 9 2 BACKPRESSURE COMPENSATION The air flow delivery of a sup...

Страница 32: ...nel when the two pieces are slid together Side extension channels are joined together by a thumbnut and splice angle arrangement The individual frame pieces are stamped with a frame number as shown ci...

Страница 33: ...added to both the top and bottom of the support dowels The dowel end pins are to be inserted into the frame corner brackets as shown in Figure 10 8 Each frame piece has been labeled to indicate posit...

Страница 34: ...ADM 870 05 23 08 30 FIGURE 10 2 FLOWHOOD IN CASE FIGURE 10 1 FRAME STORAGE...

Страница 35: ...ADM 870 05 23 08 31 FIGURE 10 3 FLOWHOOD ASSEMBLY...

Страница 36: ...ADM 870 05 23 08 32 FIGURE 10 4 2X2 FRAME ASSEMBLY FIGURE 10 5 1X4 FRAME ASSEMBLY FIGURE 10 6 2X4 FRAME ASSEMBLY...

Страница 37: ...ADM 870 05 23 08 33 FIGURE 10 7 1X5 FRAME ASSEMBLY FIGURE 10 8 3X3 FRAME ASSEMBLY...

Страница 38: ...her at any convenient point on the outside of the base Set the flaps control knob to the open position Press the READ key or the thumbswitch on the FlowHood handle and the reading will be displayed di...

Страница 39: ...with the flaps open as described above Close the flaps when the meter displays CALC When the first reading is displayed press the READ key to take the second backpressure compensated reading If necess...

Страница 40: ...in the exhaust hood It is very important that a kitchen system balance be confirmed by measurements for air flow room pressure and actual makeup air while the cooking devices are at their normal work...

Страница 41: ...use an extreme concentration of air velocity on portions of the flow sensing grid The FlowHood readings will be inaccurate under such conditions Reading accuracy is improved by adjusting the register...

Страница 42: ...the density effects of relative humidity using the formula shown below This calculation requires the use of the vapor pressure which must be looked up on a Dew Point Vapor Pressure conversion chart T...

Страница 43: ...n with the ADM 870 AirData Multimeter may be compared with hot wire anemometer readings if the hot wire readings are corrected for local air density conditions to obtain true air velocity results The...

Страница 44: ...ure source must be self replenishing such as a pressure drop across a needle valve orifice or orifice plate 14 3 ABSOLUTE PRESSURE FUNCTION Confirm the accuracy of the absolute pressure correction by...

Страница 45: ...obe is not used for either the flow or velocity measurement it must not be used for the other type of measurement In this case both sets of measurements will be calculated using standard 70 F conditio...

Страница 46: ...anything inside the meter will void the warranty and may cause serious damage to the sensitive electronic components The AirData Multimeter should be returned to the factory at least every two years...

Страница 47: ...tional cushioning to fill the carton Do not use spray foam Spray foam can damage the meter and it is possible to lose a meter in a chunk of spray foam Ship directly to Shortridge Instruments Inc 7855...

Страница 48: ...gular charges will apply This warranty shall be effective only if use and maintenance is in accordance with our instructions and written notice of a defect is given to the Seller within such period TH...

Страница 49: ...for the density effect of variations in temperature Record the AirData Multimeter temperature and absolute barometric pressure reading at the beginning and end of each probe test range These readings...

Страница 50: ...et up with different pressures of precisely the desired test differential pressure These testers must be carefully adjusted to an accurate equal base pressure of approximately 3 psi before adding weig...

Страница 51: ...the upper half positive 6 cells of the battery set by placing your DC voltmeter common ground probe at the left center battery clip and the other probe at the upper right corner battery clip The volt...

Страница 52: ...d above The temperature during storage should be less than 95 F The batteries will function best after storage if reconditioned by performing two full cycles of 24 hour charge and full discharge follo...

Страница 53: ...e IM 870 1 Instruction manual ADM 870 8400 FLOWHOOD 844 1 FlowHood base assembly with flaps 845 1 Flow sensing grid 826 1 Top support assembly 842 1 Pushbutton handle assembly 827 4 Support dowels 838...

Страница 54: ...set complete 3 x 3 TOP 91 4 x 91 4 cm 141 4 Side channel type 1 336 4 Side channel type 6 3x3F 1 Complete frame assembly all of above 3x3S 1 Cloth skirt 36 x 36 827E 8 Support dowel extenders 3x3T 1...

Страница 55: ...nnn 7 chemical exhaust hoods 19 clean room filter outlets 19 CLEAR 4 9 12 clear memory 13 constant volume controllers 36 correction factors 18 38 default air density correction 18 DENS 4 density corre...

Страница 56: ...all 6 RECHARGE 10 relative humidity 38 relative humidity correction 38 replacement parts list 49 50 reset switch 6 response time 2 safety cabinets 19 safety notice 34 sea level conditions 22 Sheet Met...

Отзывы: