9/30/2011
20
Troubleshooting
The Self-Help Guide (Table 1) is intended to assist in the location of a fault that
may occur in a certain part of the Powerchair.
If after checking the fault from the table below and the fault is still showing, do not
use the Powerchair. Turn off the power and consult your provider immediately.
Table 1
BATTERY GAUGE
POSSIBLE FAULT
10 Bars flash
An excessive voltage has been applied to the control
system. This is usually caused by a poor battery
connection.
9 Bars flash
The parking brakes have a bad connection. Check the
parking brake and motor connections.
8 Bars flash
A control system fault is indicated. Make sure that all
connections are secure.
7 Bars flash
A joystick fault is indicated. Make sure that the joystick is in
the center position before switching on the control system.
6 Bars flash
Battery charger connected.
5 Bars flash
Right-hand motor wiring fault.
4 Bars flash
Right-hand motor disconnected.
3 Bars flash
Left-hand motor wiring fault.
2 Bars flash
Left-hand motor disconnected.
1 Bar flashes
Low battery voltage.
Battery/ Charging Troubleshooting
– If the charger indicator light goes to green within a few minutes after being plug
in and doesn’t charge the batteries the batteries may be over discharged and
may require additional service or replacement. Contact your provider for
assistance.
If the charger indicator light goes immediately to green when plugged in to
charge, the problem is likely a blown fuse, popped circuit breaker (push reset
button) or similar. Contact your provider for assistance. If the charger does not
come on at all when plugged in, check the charger fuse if applicable, then call
your provider
.