7.4 Column Cleaning (Regeneration) 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning and/or regenerating the column. When washing and/or regenerate the 
column, let the washing solution flow from the column outlet go directly into the waste container and not let the 
solution go through the detector. 

 

<Cleaning Method> 
Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 
introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.   

 

<Regeneration Method> 

 

Presence of cations other than lead ion in the sample may replace the lead ion attached to the sulfo functional 
group. This can change the separation patterns and peak shapes. Following regeneration procedure reattaches 
the lead ions and may resume the column performance.   

 

Set the column temperature between 70 - 95 °C and introduce water at 0.5 mL/min flow rate.   
Inject 50 

μ

L of 0.2-M aqueous lead nitrate solution several times.   

 

 Caution! 

 

 

· Please be careful when disposing the eluate containing lead nitrate solution. 

 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 

9. Column Inspection

 

Inspection method is described in the Certificate of Analysis (CoA).   
Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations.   

 

(1) Theoretical Plate Number (N) 

 

 

(2) Asymmetry Factor (Fas) 

 

 

 
 
 
 

Содержание SUGAR SP0810 8C

Страница 1: ... Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex SUGAR SP0810 8C is a high performance chromatography column which is specifically designed for the separations and analysis of sugars A strong acid cation exchange type packing gel prepared from a rigid resin of styrene divinylbenzene copolymer provides a mixed separation mode The analytes are ...

Страница 2: ...have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elution patterns and or damage the column Note Use of on line degasser is recommended 6 Sample Preparation 1 If possible use the eluent for analysis to dissolve or dilute samples If this is difficult use a solvent wh...

Страница 3: ...the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired 4 At the end of analysis reduce the flow rate to 0 2 0 3 mL min and turn the column oven off Make sure that the column oven temperature is at room temperature before turning off the pump Caution Verify that there is n...

Страница 4: ...he lead ion attached to the sulfo functional group This can change the separation patterns and peak shapes Following regeneration procedure reattaches the lead ions and may resume the column performance Set the column temperature between 70 95 C and introduce water at 0 5 mL min flow rate Inject 50 μL of 0 2 M aqueous lead nitrate solution several times Caution Please be careful when disposing the...

Страница 5: ...d surface 3 When disposing the column please follow a proper waste disposal method specified by your local regulations Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: