-3 Tube Connections
3 Tube Connections
3 Tube Connections
3 Tube Connections
When connecting this detector to a high performance liquid chromatograph through
tubes, wash the chromatograph and columns thoroughly with the eluent in use and
make sure that the eluent is clean before connecting the tube.
Connect the tubes in accordance with the following procedures :
1)
Connect the eluent inlet joint
to the outlet of the column to be used
through the provided IN TUBE.
When fastening the stainless steel set screw of the IN TUBE, grip the eluent
inlet joint
with a spanner.
2)
Connect the supplied OUT TUBE to the eluent outlet joint
.
Do not connect any tube other than the provided OUT TUBE to the
Eluent outlet joint
of the detector.
Put the exit end of the OUT TUBE in a waste liquid receptacle and do
not apply back pressure.
When connecting this detector to other detectors in series, connect
this detector at the end.
□