7.4 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column. When washing the column, let the washing solution flow 
from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1)

 

Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below example for washing the adsorbing components. For an efficient cleaning, reverse the flow 

direction. 

 

Introduce 50 mL of acetonitrile or methanol at 0.5 mL/min.   

 

8. Column Storage

 

Remove the column from the system keeping the solvent used for the analysis. Securely tighten the end caps 
and store the column at a location with stable temperature (a cool and dark space is recommended). 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 

9. Column Inspection

 

Inspection method is described in the Certificate of Analysis (CoA).   
Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations.   

 

(1) Theoretical Plate Number (N) 

 

 

(2) Asymmetry Factor (Fas) 

 

 

 

10. Additional Warnings 

(1) Do not remove end fittings. 

 

(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. 

 

Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications.   
For additional assistance, contact the distributor from whom you purchased the column or contact your regional 
Shodex

 support office (https://www.shodex.com/en/support_office/list/). 

 

 
 

Содержание ORpak CDBS-453

Страница 1: ...outside of the column for any damage 2 Please check if product name and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex ORpak CDBS 453 is a chiral separation column filled with spherical porous particles of silica modified with β cyclodextrin group Optical...

Страница 2: ...bottle Use HPLC grade organic solvent whenever possible Solvents left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has been contaminated Always use freshly prepared solvents Solvents stored for a long time may have changed their compositions and may influence elution patterns and or damage the column Note Use of on line degasser is r...

Страница 3: ... start the system at less than 0 2 mL min A rapid increase in pressure can damage the column If the column was used at an elevated temperature lower the flow rate to less than 0 2 mL min at the end of analysis Then turn off the column oven and let the column temperature return to room temperature before stopping the pump If the pump was stopped while the eluent inside the column was still hot as t...

Страница 4: ...the flow direction Introduce 50 mL of acetonitrile or methanol at 0 5 mL min 8 Column Storage Remove the column from the system keeping the solvent used for the analysis Securely tighten the end caps and store the column at a location with stable temperature a cool and dark space is recommended Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Inspectio...

Страница 5: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: