Shodex ODP2 HP-2B Скачать руководство пользователя страница 2

ver.211124E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

4.

 

Usable Conditions 

 

Usable solvents are listed below. 
(1) A mixture of water and acetonitrile or water and methanol of any ratio can be used. Up to 100 % water, 

acetonitrile, or methanol can be used. For the analysis of aromatic compounds, if the theoretical plate 
number obtained was significantly lower than expected when using methanol, try using acetonitrile instead.   

 

(2) Up to 100 % ethanol, isopropanol (IPA), and tetrahydrofuran (THF) can be used. 

 

 

(3)

 

Various buffer and aqueous salt solutions can be used instead of water. They include buffers such as 
phosphate, acetate, and Tris and salts such as sodium chloride and potassium chloride. Please keep total 
concentration of salts under 0.1 M. Their recommended concentration ranges are 1 - 50 mM. Acids like 
trifluoroacetic acid (TFA), formic acid, and acetic acid can also be used.   

 

(4)

 

Up to 2-M ammonium sulfate can be used.   

 

 

 

Attention!

   

· Use the column within above stated flow rate, pressure, and temperature ranges. Using 

the column outside the given range may damage the column and lower its performance.   

 

· When using a mixture of buffer or salt solution and organic solvent, make sure there is no 

precipitation of salt. 

 

· When using highly corrosive salts such as sodium chloride, wash out the salts at the end 

of analysis. The metal parts of the devices and/or the columns may rust.

 

 

 

·

 

Column pressure is influenced by the eluent composition, flow rate, and column  

temperature. When changing the eluent compositions, adjust the flow rate and column 
temperature so that the column pressure remains below the usable maximum pressure.   

 

5. Eluent Preparation 

(1) Degas the eluent fully to prevent the formation of air bubbles.   

 

(2) Presence of small debris or insoluble substances may result in deterioration of the column and/or they appear 

as noise on the chromatograms. Filter the eluent with a 0.45-

μ

m disposable filter to prevent the problems. 

 

 

Attention!

   

· Whenever water is required, use ultra-pure water freshly generated by a water purification 

system or water from a newly opened HPLC grade distilled water bottle. Use HPLC grade 
organic solvent whenever possible. Solvents left in an opened bottle for a long time should 
not be used. The content may have been changed, absorbed moisture, or has been 
contaminated. 

 

· Always use freshly prepared solvents. Solvents stored for a long time may have changed 

their compositions and may influence elution patterns and/or damage the column.

   

 

 

       

Note  

 

 

· Use of on-line degasser is recommended. 

 

6. Sample Preparation 

(1) If possible, use the eluent for analysis to dissolve or dilute samples. If this is difficult, use a solvent which 

has a composition that is as close as possible to the eluent's composition, but which fully dissolves or dilutes 
the sample. When gradient elution is used, it is recommended to use the initial eluent to prepare the sample.     

 

(2) Filter the sample solution using disposable 0.45-

μ

m filter to prevent the column from clogging or deteriorating. 

 

(3) Recommended sample injection volume is less than 5 

μ

L per column. 

 

 

       

Note   

 

· Use of guard column is recommended to protect the analytical column. 

 
 
 

Product Name 

Flow Rate (mL/min) 

Maximum 

Pressure (MPa) 

(Per Column) 

pH 

Range 

Temperature 

Range (°C) 

Recommended Maximum 

ODP2 HP-2B 

0.1 - 0.2 

0.4 

12 

3 - 12 

20 - 60 

ODP2 HP-2D 

15 

ODP2 HPG-2A 

- - - 

Содержание ODP2 HP-2B

Страница 1: ... 1 Introduction Thank you for purchasing the Shodex product Shodex ODP2 HP series is polymer based reversed phase chromatography columns which retain hydrophilic compounds stronger than ordinarily ODS columns do The small pore size of the packing material excludes proteins fast and this makes the columns suitable for separating small molecular weight components such as drugs even in the presence o...

Страница 2: ...t the column pressure remains below the usable maximum pressure 5 Eluent Preparation 1 Degas the eluent fully to prevent the formation of air bubbles 2 Presence of small debris or insoluble substances may result in deterioration of the column and or they appear as noise on the chromatograms Filter the eluent with a 0 45 μm disposable filter to prevent the problems Attention Whenever water is requi...

Страница 3: ...nd fitting Presence of an extra space will let the sample to spread out and may result in wide peaks 3 Set the initial flow rate at less than 0 1 mL min and start the system If using the column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired Caution Verify that there is no so...

Страница 4: ...the column outlet go directly into the waste container and not let the solution go through the detector Cleaning Method 1 Insoluble components that block the column inlet may be removed by reversing the flow direction i e introducing the eluent from the column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficien...

Страница 5: ... Solvent Exchange for how to replace the eluent Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Inspection method is described in the Certificate of Analysis CoA Theoretical Plate Number N and Asymmetry Factor Fas were calculated using the below equations 1 Theoretical Plate Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fi...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: