Shodex IEC SP-FT 4A Скачать руководство пользователя страница 4

ver.211110E

 

SHOWA DENKO K.K. (https://www.shodex.com/)

 

·

 

Analytes are retained for very short time when using an ultra-rapid analysis and this makes 

their peak shapes sharp. Slow detector response results in broader peaks and this may 
lead to bad peak separations than actual. Please optimize the detector response setting 
by checking the peak separations. Also, slow data-processor sampling rate results in less 
data point collection and this may lead to bad peak shapes. Please set the data-
processor’s sampling rate to obtain more than 20 data points for each peak width. 

 

       

Note  

 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 

7.3 Solvent Exchange 

(1) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(2)

 

When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 
use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

(3) When using a gradient method, changes in the eluent compositions may increase the column backpressure. 

Adjust the flow rate and column temperature so that the column backpressure remains below the usable 
maximum pressure.

 

 

7.4 Column Cleaning 

Problems in peak shapes and elution time changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column.   
Make sure to clean the guard and the analytical columns separately. When washing the column, let the washing 
solution flow from the column outlet go directly into the waste container and not let the solution go through the 
detector. 

 

<Cleaning Method> 
Follow below cleaning steps for adsorbing components. For an efficient cleaning, use the flow rate less than   
0.5 mL/min. Introduce 5 to 10 times column volume of the washing solvent. 

 

Method 1: Adsorption of proteins   
Introduce an eluent with high salt concentration (1.0 M) or 0.1 % nonionic surfactants.   

 

Method 2: Adsorption of hydrophobic compounds   
Introduce a mixture of solvent containing 50-mM eluent (buffer or aqueous salt solution) and 10 - 20 % (v/v) 
acetonitrile.  
 

8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание IEC SP-FT 4A

Страница 1: ...and serial number written on the column package column name tag and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex IEC SP FT 4A is a strong cation exchange column filled with non porous hydrophilic polymer based gel modified with sulfopropyl group Its small hydrophobic adsorption makes the column suitable for analyzin...

Страница 2: ...lt When using highly corrosive salts such as sodium chloride wash out the salts at the end of analysis The metal parts of the devices may rust Column pressure is influenced by the eluent composition flow rate and column temperature When changing the eluent compositions adjust the flow rate and column temperature so that the column pressure remains below the usable maximum pressure 5 Eluent Prepara...

Страница 3: ... 2 Column Installation 1 Connect the column to LC system by following the flow direction arrow indicated on the column name tag If guard column is used position the guard column in front before the inlet of the analytical column 2 Make sure to insert the tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the col...

Страница 4: ...usable maximum pressure 7 4 Column Cleaning Problems in peak shapes and elution time changes or elevated column pressure etc are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column These problems may be resolved by cleaning the column Make sure to clean the guard and the analytical columns separately When washing the column let the...

Страница 5: ...late Number N 2 Asymmetry Factor Fas 10 Additional Warnings 1 Do not remove end fittings 2 Do not make a strong impact on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support of...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: