[Guard filter (HK-G): Installation] 

(1) At the first time use, connect the flow line to the guard filter inlet alone (do not connect to the analytical 

column), set the flow rate at 0.3 mL/min and start the system. Leave it for about 1 minute.   

 

(2) Stop the system and connect the outlet of the guard filter to the inlet of the analytical column, tighten by 

hand. Place wrenches on the parts indicated by 

  and tighten additional 30 - 60

o

. After connecting the 

guard filter and the analytical column, they can be kept connected together during the storage time.   

 

 

 

 

Attention!

   

· Tightening too much with wrenches can damages the thread.     

 

· Do not place a wrench on the part indicated by

 

 

to tighten or remove the guard filter. 

This may loosen the end fitting of the analytical column and results in solvent leak. 

 

 

[Guard filter (HK-G): Filter replacement] 

(1) Place wrenches on the parts indicated by 

  and untighten the connection. Detach the guard filter from 

the analytical column. Cap the inlet of the analytical column. 

 

(2) Use wrenches to untighten the guard filter. Replace the old filter with a new filter. 

 

(3) Tighten the holder by hands and then use wrenches to tighten additional 180

o

.  

 

(4) Refer to the steps described in “Guard filter (HK-G): Installation” to connect the guard filter to the analytical 

column.  

 

 

Attention!

 

 

· Make sure to use an official replacement filter (product code: F6700100). Product 

performance cannot be guaranteed if unofficial filter was used.   

 

7.3 Solvent Exchange 

(1) When replacing the solvent, set the column temperature at 40 - 60 °C and start the system at less than     

0.3 mL/min. Recommended solvent volume to introduce at each step is 3 to 5 times of the column volume. 

 

(2) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column. 

 

(3) When replacing the current solvent with a miscible solvent, first introduce 1:1 mixture of the current solvent 

and the new solvent, and then replace it with 100 % of the new solvent. 

 

e.g. When replacing THF to DMF, first run 1:1 mixture of THF and DMF and then introduce 100 % DMF. 

 

(4) When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first 

use a solvent that is miscible/soluble to both eluents, and then replace it with the new solvent. 

 

e.g. When replacing THF to DMF with LiBr, first follow step (3) above, and then replace 100 % DMF to DMF 

with LiBr.   

 

 

Attention!

   

· Frequent solvent replacement deteriorates the column, and thus not recommended.

     

 

 

 
 
 

 

 
 
 
 

Содержание GPC HK-400 Series

Страница 1: ...oduct Shodex GPC HK 400 series is an ultra rapid analysis type size exclusion chromatography column used with organic solvents The column series is suitable for the separations of organic compounds and or oligomers and molecular weight distribution analysis of high molecular weight polymers The columns with different exclusion limits are available Please select columns with target molecular weight...

Страница 2: ...osion when peroxide is concentrated An extra caution should be taken when concentrating the collected sample solution after separation Attention Use the column within above stated flow rate pressure and temperature ranges Using the column outside the given range may damage the column and lower its performance THF without BHT easily forms peroxides and this causes unstable baseline If necessary to ...

Страница 3: ...ent for analysis to dissolve or dilute samples If this is difficult use a solvent which has a composition that is as close as possible to the eluent s composition but which fully dissolves or dilutes the sample 2 Filter the sample solution using disposable 0 45 μm filter to prevent the column from clogging or deteriorating 3 To prepare a sample with molecular weight larger than 1 000 000 first all...

Страница 4: ...e column at an elevated temperature keep a low flow rate until the temperature of the column reaches to the set temperature and then gradually increase the flow rate to the desired 4 Multiple number of columns can be connected in series When connecting multiple columns with different exclusion limits set the column with higher exclusion limit at the upper stream position When connecting HK 404L us...

Страница 5: ...y hands and then use wrenches to tighten additional 180o 4 Refer to the steps described in Guard filter HK G Installation to connect the guard filter to the analytical column Attention Make sure to use an official replacement filter product code F6700100 Product performance cannot be guaranteed if unofficial filter was used 7 3 Solvent Exchange 1 When replacing the solvent set the column temperatu...

Страница 6: ...tainer and not let the solution go through the detector Cleaning Method Insoluble components that block the column inlet may be removed by reversing the flow direction i e introducing the eluent from the column outlet with flow rate at less than half of the recommended flow rate 8 Column Storage Remove the column from the system keeping the solvent used for the analysis Securely tighten the end ca...

Страница 7: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: