7. Column Usage Procedure 

 

7.1 HPLC System Preparation 

Wash entire LC system prior to the column installation, including all flow-lines and sample loop by switching the 
valve, and then replace the washing solution with the eluent to be used. If desired new eluent has low 
miscibility/solubility to the eluent of previous analysis, first use the eluent that is miscible/soluble to both eluents, 
and then replace it with the desired eluent.   

 

e.g. When replacing chloroform to water/acetonitrile, first run methanol and then replace it with water/acetonitrile. 

 

e.g. When replacing highly concentrated buffer solution or salt solution to water/acetonitrile, first run water and 

then replace it with water/acetonitrile. 

 

 Attention! 

 

· If the eluent left in the system is not compatible with the column to be used, it may damage 

the column. 

 

 

     · A drastic change in the eluent compositions may remove substances adsorbed on the 

system and they may enter and deteriorate the column. 

 

7.2 Column Installation

 

 

(1) Connect the column to LC system by following the “flow direction arrow” (

) indicated on the column 

adhesive label. If guard column is used, position the guard column in front (before the inlet) of the analytical 
column. 

 

(2) Make sure to insert the tubing all the way to the end fitting and secure it with the male nut. It is important that 

there is no extra space between the tubing and the column side of the end fitting. Presence of an extra space 
will let the sample to spread out and may result in wide peaks. 

 

 

 

(3) Set the initial flow rate at 2.5 mL/min and start the system. If using the column at an elevated temperature, 

keep a low flow rate until the temperature of the column reaches to the set temperature, and then gradually 
increase the flow rate to the desired.   

 

 Caution! 

 

 

· Verify that there is no solvent leak. It may cause electronic leakage, rust, and/or chemical 

injury. 

 

 

Attention!

   

· Make sure not to let air bubbles enter the column while installing the column. The air 

bubbles may damage the column.   

 

· When restarting the system after column installation or after holding the eluent flow, start 

the system at less than 2.5 mL/min. A rapid increase in pressure can damage the column.   

 

· If the column was used at an elevated temperature, lower the flow rate to less than      

2.5 mL/min at the end of analysis. Then, turn off the column oven, and let the column 
temperature return to room temperature before stopping the pump. If the pump was 
stopped while the eluent inside the column was still hot, as the eluent temperature 
decreases, its volume also decreases. This may result in creating an empty space in the 
column and deteriorates the column. 

 

       

Note  

 

 

· It is recommended to set the pump limiter to avoid exceeding the maximum pressure.   

 
 
 
 
 
 
 
 

Содержание Asahipak NH2P-90 20F

Страница 1: ...ckage and outside of the column for any damage 2 Please check if product name and serial number written on the column package column adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak NH2P series is a polymer based chromatography column modified with amino group The column is suitable for the a...

Страница 2: ...h a 0 45 μm disposable filter to prevent the problems Attention Whenever water is required use ultra pure water freshly generated by a water purification system or water from a newly opened HPLC grade distilled water bottle Use HPLC grade organic solvent whenever possible Solvents left in an opened bottle for a long time should not be used The content may have been changed absorbed moisture or has...

Страница 3: ...tubing all the way to the end fitting and secure it with the male nut It is important that there is no extra space between the tubing and the column side of the end fitting Presence of an extra space will let the sample to spread out and may result in wide peaks 3 Set the initial flow rate at 2 5 mL min and start the system If using the column at an elevated temperature keep a low flow rate until ...

Страница 4: ...solution Please refer to 7 4 Column Cleaning Method 1 7 4 Column Cleaning Problems in peak shapes and elution time changes or elevated column pressure etc are often caused by insoluble or adsorbing components present in the eluent and reagents being deposited inside the column These problems may be resolved by cleaning the column If a guard column is used with the analytical column first remove th...

Страница 5: ...and store the column at a location with stable temperature a cool and dark space is recommended Refer to section 7 3 Solvent Exchange for how to replace the eluent Attention Never allow inside of the column to dry It can damage the column 9 Column Inspection Inspection method is described in the Certificate of Analysis CoA Theoretical Plate Number N and Asymmetry Factor Fas were calculated using t...

Страница 6: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: