7.4 Column Cleaning 

Problems in peak shapes and elution timing changes or elevated column pressure etc. are often caused by 
insoluble or adsorbing components present in the eluent and reagents being deposited inside the column. These 
problems may be resolved by cleaning the column.   
If a guard column is used with the analytical column, first remove the guard column and check the performance 
of the analytical column alone. If the problem is solved, most likely the cause is from the guard column. In this 
case, clean the guard column.   
If the problem is not solved by removing the guard column, clean both guard and analytical columns. Make sure 
to clean the guard and the analytical columns separately. When washing the column, let the washing solution 
flow from the column outlet go directly into the waste container and not let the solution go through the detector. 

 

<Cleaning Method> 
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e., 

introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate. 

 

(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction 

and use the flow rate at less than 3.0 mL/min. Introduce 5 to 10 times column volume of the washing solvent. 

 

Method 1: Adsorption of hydrophilic compounds (when using aqueous eluent) 

Introduce the eluent with a high polar organic solvent (acetonitrile or methanol) added. 

 

Method 2: Adsorption of ionic compounds 

  Introduce the eluent with a higher salt concentration.     

 

 

Attention!

   

· Keep the organic solvent and salt within the concentrations stated in 4.2 Usable Solvents. 

 
8. Column Storage

 

Remove the column from the system after replacing the in-column solvent with the shipping solvent. Securely 
tighten the end caps and store the column at a location with stable temperature (a cool and dark space is 
recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. 

 

 

Attention!

   

· Never allow inside of the column to dry. It can damage the column.     

 

9. Column Inspection

 

Inspection method is described in the Certificate of Analysis (CoA).   
Theoretical Plate Number (N) and Asymmetry Factor (Fas) were calculated using the below equations.   

 

(1) Theoretical Plate Number (N) 

 

 

(2) Asymmetry Factor (Fas) 

 

 

Содержание Asahipak GS-310 20G

Страница 1: ...check if product name and serial number written on the column package column adhesive label and enclosed Certificate of Analysis CoA are matching and correct 1 Introduction Thank you for purchasing the Shodex product Shodex Asahipak GS 310 20G and GS 510 20G are polymer based size exclusion chromatography columns They work well with both aqueous and organic solvents 2 Column Components 3 Column Sp...

Страница 2: ...ent It is recommended to dedicate the column for this specific use 4 Surfactants such as SDS and Brij 35 can also be added to the eluent However since surfactants tend to remain on the column solvent replacement after their use takes a longer time than replacement from general solvents The replacement time can be shorten by using 30 50 v v methanol Attention Use the column within above stated flow...

Страница 3: ... clogging or deteriorating 3 Recommended injection volume is less than 1 mL per column 4 Viscosity of high molecular weight compound is largely influenced by its molecular weight and concentration Samples with high viscosity cause peak broadening and elution delay and this makes it difficult to obtain their accurate molecular weight distributions In general the larger the molecular weight of the c...

Страница 4: ... at less than 3 0 mL min A rapid increase in pressure can damage the column If the column was used at an elevated temperature lower the flow rate to less than 3 0 mL min at the end of analysis Then turn off the column oven and let the column temperature return to room temperature before stopping the pump If the pump was stopped while the eluent inside the column was still hot as the eluent tempera...

Страница 5: ... column outlet with flow rate at less than half of the recommended flow rate 2 Follow below cleaning steps for adsorbing components For an efficient cleaning reverse the flow direction and use the flow rate at less than 3 0 mL min Introduce 5 to 10 times column volume of the washing solvent Method 1 Adsorption of hydrophilic compounds when using aqueous eluent Introduce the eluent with a high pola...

Страница 6: ...ct on the column Do not drop or hit the column on a hard surface Refer to the Shodex website https www shodex com for product details and their applications For additional assistance contact the distributor from whom you purchased the column or contact your regional Shodex support office https www shodex com en support_office list ...

Страница 7: ...anyone other than person or firm designated by Showa Denko K K 3 3 If the Shodex product is disposed of 3 4 If the Shodex product is resold by the user without giving prior written notice to Showa Denko K K 3 5 If the performance of the Shodex product is not conform to the specification of the attached CERTIFICATE OF ANALYSIS due to any of the reasons below a Computer virus b Impurities contained ...

Отзывы: