16
4.4 Interface
Displayed when [6. I.F.] is [1-4].
Item Set
Value
Description
0
Stop output
1
Output continuously at all times
2
Output continuously if stable (Stop output if unstable)
3
Output once when the Print key is pressed (whether the balance
is stable or unstable).
4
Output once when the balance is stable. Output when a substance
is lifted to cause the display to indicate a value below 0, and then
another substance is put on to make the balance stable.
5
Output once when the balance is stable. Stop output when
unstable. Output once when the balance is stabilized again even if
it is not reloaded (the output includes zero).
6
Output once when the balance is stable. Output continuously
when unstable. Output is stopped after a single output when the
balance is stable even if it is not reloaded.
7
Output once when the Print key is pressed if the balance is stable.
A
Output once immediately after a specified interval.
Output Control
61. o.c.
*1
b
Output once immediately after a specified interval if the balance is
stable.
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
Baud Rate
62. b.L.
5
19200 bps
0
0
None
1 Odd
Parity
63. PA.
2
2
Even
Displayed when [6. I.F.] is set to [2to4].
7
7 bits
Data Length
64. d.L.
8
8 bits
1
1 bit
Stop Bits
65. St.
2
2 bits
Displayed when [6. I.F.] is set to [3,4].
0
Embed 0 (30H).
Unused High Order
Digits
66. u.n.
1
Embed space (20H).
1
Use A00 and Exx.
Response Commands 67. r.ES.
2
Use ACK and NAK.
A star ( ) denotes a factory setting.
*1 To connect this instrument to a printer, select one numeral or letter from among 0, 2, 4, 5, 7, and b.
The data output interval is 0.1 to 1 second for continuous output. (The interval varies with the
measurement conditions, etc.)