- 21 -
Force limit
The stand controls to give the force less than the "Force limit".
If the force limit is detected, the stand will stop immediately. Then the stand executes the next step.
Detect object
The parameter is the threshold force level for detecting the object.
When the setting force of "Detect object" parameter is detected, the stand will stop immediately. Next
the stand moves to opposite direction until detecting "Zero force".
The recommended "Detect object" that the user should program into the stand is more than 0.2% of
the full scale of the mounted force gauge, i.e., if the programmer is using an FGV-200H force gauge,
the programmed "Detect object" should be 0.4lb (200lb x 0.2%).
5.5. Parameter setting mode
The relations function is set to the entire stand.
Contents
The parameters are shown in the below table.
Items
Description
Range
Sign of Push direction
The sign at the position of the direction of
Push is set.
If it sets "-" and the stand moves to Push, the
display of position is decrement.
[space], -
Acceleration
The acceleration of the motor of the stand is
set.
The acceleration means the slope from
motor starting to reach to the setting speed.
If the parameter sets maximum 100, the
motor moves quickly.
1 to 100 (step 1)
Volume of buzzer
The volume of the buzzer is adjusted.
OFF, Low, Middle, High
Upper limit
Lower limit
Start point
Detect point
Standard point
Point 1
Point 4
Point 5
Point 3
Point 2
Example
Point 4
Point 5
Detect
the force limit