3
2. Name and Function of Components
Read These
Pages
Before
Installation
2.2 Key Panel and Operation
Functions of the keys
*1 This key is used to set values when percent (%), number (PCS), solid specific gravity (
T
d), or liquid specific
gravity (d) are displayed.
*2 When a Pretare value is set, zero is not displayed and [- Pretare value] is displayed.
*3 Units other than “g” must be set up before they can be used for measurement. Only gram (g), percent (%), and
piece counting (PCS) are set-up before shipment. To set up other units or specific gravity measurement, refer
to section
*4 When the unit is set to 10d, the resolution of the minimum display is decreased by one decimal place.
Key
During Weighing
Press Once and Release
Press and Hold for About 3 Seconds
[POWER]
Switches between the operation and standby
modes.
Exits the application function and returns to
the mass display.
[CAL]
Enters span calibration or menu item selection.
(*1)
Displays the last menu item that was set.
(Last menu recall)
[O/T]
Tares the balance. (Displays zero.) (*2)
Displays the Pretare value.
[UNIT]
Changes the weighing unit or selects specific
gravity measurement. (*3)
Switches between the 1d and 10d display. (*4)
[PRINT]
Sends the displayed value to a peripheral device.
Sends the date and time to a peripheral
device.
Key
During Menu Item Selection
Press Once and Release
Press and Hold for About 3 Seconds
[POWER]
Returns to the previous menu level
Returns to the mass display.
[CAL]
Moves to the next menu item.
Displays the last menu item that was set.
(Last Menu Recall)
[O/T]
Selects or sets the currently displayed menu item,
or enter into the displayed menu.
No operation.
[UNIT]
Increases the numeric value of the blinking digit by 1.
No operation.
[PRINT]
Moves to the next digit during numeric value entry.
No operation.
[POWER]
key
[CAL]
key
[O/T]
key
[UNIT]
key
[PRINT]
key