5-22
11 Press in the handle of the condenser lens unit to place the lens at its original
position.
Caution
Do not press in the condenser lens unit without the condenser lens
or without the O-ring (1A P34).
It may break the vacuum inside the spectrometer and contaminated
water in the spark stand or bubbler may enter the spectrometer. As a
result, the optical units may be contaminated, causing permanent
damage to the instrument.
12 Connect the vinyl hose for exhausting argon gas for the spark stand to the stand
following step 9 described in "5.3 Maintaining the spark stand".