Shimadzu Bransist Alexa Скачать руководство пользователя страница 34

34

 

Controlling Image Playback

IVR NEO Joystick

Keyboard

Control Bar

 

Playback

Control Bar

Keyboard

IVR NEO Joystick

Function

Play/show next loop/image.

Play/show previous loop/image.

 or 

(Press button)

Pause

Advance to next frame. (when paused)

Return to previous frame. (when paused)

 ( Press more 

than 2 seconds)

(Press button)

Resume forward play.  
(same speed as acquisition)

 ( Press more 

than 2 seconds)

Resume backward play.  
(same speed as acquisition)

No.

Button

Function

1

Adjust the loop playback speed.

Right-click:  + (increase)
Left-click:  - (decrease)
Click both mouse buttons simultaneously: Return to default

2

Change loop playback mode.

Play Single Loop with Synchronize:
Plays back selected single loop until a new loop is received. Then, the 
new loop is selected and played back continuously.
Play Single Loop:
Plays back the selected loop continuously.

Содержание Bransist Alexa

Страница 1: ...Guide BRANSIST alexa Digital Angiography System M517 E190D Digital Angiography System BRANSIST alexa Read this manual thoroughly before you use the product Keep this manual for future reference Operat...

Страница 2: ...a potentially hazardous situation which if not avoided may result in minor to moderate injury or equipment damage NOTE Emphasizes additional information that is provided to ensure the proper use of t...

Страница 3: ...Change Settings of Injection Delay 31 Recording Fluoroscopy Image 32 Adjusting Image 33 Adding Deleting and Displaying Reference Images 35 Related Operation for Study 36 Measuring and Analyzing 37 Ope...

Страница 4: ...Tube Collimator IVR NEO Catheterization Table CyberGrip or Cyber Console Local Console Foot Switch Examination Room Acquisition Monitor Review Monitor C arm Indicator C arm 1 System Components Examin...

Страница 5: ...r side of the grip C arm Rotation Selector Switch Direct Memory Switches C arm Travel Selector Switch FPD SID Switch Cyber Console Field of View Switch Center Reset Switch C arm Rotation Travel FPD SI...

Страница 6: ...h Direct Memory Switches Operate the C leaf of the Collimator by moving up and down Operates the H V Collimation The Collimator opens up when this is pressed down Operates the Compensation Filter in t...

Страница 7: ...Direct Memory Switches Operate the C leaf of the Collimator by moving up and down Operates the H V Collimation Operates the Compensation Filter in the Collimator Leaves Oper Switch The Collimator ope...

Страница 8: ...ct a button from and 2 Select the location of the C arm from this area 3 Press and hold to move the C arm to memory position Type B 1 Select a button from 2 Select the location of the C arm from this...

Страница 9: ...ype A 1 Select a button from and 2 Select the location of the C arm from this area 3 Press and hold to move the C arm to memory position Type B 1 Select a button from 2 Select the location of the C ar...

Страница 10: ...Comment ON Comment menu displayed Add Reference Image ON can add Reference Image Fluoro Record Last N Seconds mode ON can record the latest uoro loop unavailable for Direct Fluoro Record mode Recordin...

Страница 11: ...select mode LIVE ON display DSA LIVE image Linkage with C arm Positioning ON send projection angle of currently displayed image to the C arm Image Filter ON lter and display the projection angle of i...

Страница 12: ...12 Review Monitor X ray Generator Console Foot Switch Acquisition Monitor Keyboard Digital Operation Cabinet Mouse C arm Remote Console Control Room...

Страница 13: ...scopy Density Button IBS Button BH Filter Number C arm Information X ray OFF Button Setting Button Accumulated Fluoroscopy Time Reset Button Warm up Button Tool menu C arm Collision Indicator Focus Sw...

Страница 14: ...isplayed in reference mode Side Menu displayed when the mouse pointer is put on the left edge of the screen Parameters Acquisition Active Image Selector Status Messages Programs Comment Cine Control E...

Страница 15: ...Side Menu Image Viewer Probe Signal Digital Program Menu is displayed in reference mode Status Messages Programs Comment Cine Control Exit Button Modes Info Network Activity Study Information Study I...

Страница 16: ...16 Startup Be sure to perform daily inspection before patient use Daily Inspection P 21 NOTE X ray Generator Console 1 Press ON Starting up whole system Power ON Button 2 System Startup and Shutdown...

Страница 17: ...at the park position even though the uoroscopy at park position setting is set to OFF In this setting X ray exposure by the foot switches and the hand switch is not available NOTE X ray Generator Con...

Страница 18: ...18 If you need to pause warm up during procedure Press to pause warm up Press to resume Press to cancel warm up and press to con rm...

Страница 19: ...tion Press and hold the C arm Parking Switch until travel is completed C arm Type B 1 Set the C arm to the center position Press the Center Reset Switch to select the operation Then press and hold the...

Страница 20: ...is displayed 4 Con rm the message by clicking X ray Generator Console 5 Press OFF when System Display window is changed Make sure the OFF button is lit To restart the system wait 1 minute after turnin...

Страница 21: ...are no abnormal sounds LAO RAO CAUD CRAN operation of the C arm Change the position of the C arm and move the FPD forward backward Up and down operation of Catheterization Table and Tabletop Lock Swit...

Страница 22: ...the items with a dot 3 Click Patient information will appear on the upper area of the main window If loops images have not yet been recorded saved study information can be modi ed by clicking on the s...

Страница 23: ...e up down direction while pressing C arm rotates according to the direction of the grip 4 C arm Travel Move the grip in the up down direction or twist it in the left right direction while pressing C a...

Страница 24: ...on selector switch C arm rotates according to the direction of the lever C arm Rotation Selector Switch upper on the far side 4 C arm Travel Move the C arm rotation travel lever forward backward or in...

Страница 25: ...loor mount type No Switch C arm position Home position 0 Femoral position CW45 Peripheral position CW40 Movable range of the C arm rotation is LAO90 RAO 90 in the femoral position however limited to L...

Страница 26: ...ended type C arm Working Positions Switch No Switch C arm position Head position Left position Left sheltered position Right position Left diagonal position press simultaneously Right diagonal positio...

Страница 27: ...zation Table Move the table manually while pressing the unlocking switch CyberGrip Grip Switch Table Console 2 Raising of Catheterization Table Press Up switch Tabletop rises CyberGrip Table Console 3...

Страница 28: ...ppears on the screen To stop the buzzer press alarm stop button Radiography 1 Con rm the radiography program Modify the condition as needed Positioning shall be set by uoroscopy Perform uoroscopy more...

Страница 29: ...directory with mouse Selected parameter is set Fluoroscopy Radiography program presets and corresponding parameters are changeable These parameters can be changed at the digital processing system in...

Страница 30: ...setting back to normal Flip vertical Click the button again to change the setting back to normal 2 Perform uoroscopy or radiography Reversed image is acquired It is selectable during fluoroscopy or r...

Страница 31: ...n Times are set in advance for each acquisition program These values can be changed temporary during study according to the following procedure 1 Click on the review monitor side menu Acquisition Para...

Страница 32: ...beforehand NOTE Last N seconds Fluoroscopy Record 1 Perform uoroscopy 2 Click or press on IVR NEO The latest uoroscopy loop will be saved Direct Fluoroscopy Record 1 Click Make sure that selected Fluo...

Страница 33: ...click decrease Click both mouse buttons simultaneously 0 Convolution Right click increase Left click decrease Click both mouse buttons simultaneously 0 Negative positive conversion H Flip Images can b...

Страница 34: ...Resume forward play same speed as acquisition Press more than 2 seconds Resume backward play same speed as acquisition No Button Function 1 Adjust the loop playback speed Right click increase Left cl...

Страница 35: ...ay frame by frame backward according to the speed of joystick angle when paused Larger the tilt angle higher the speed Keyboard IVR NEO Function Add Reference Image Delete Reference Image Go to next l...

Страница 36: ...Function 1 New study Start examination 2 Re synchronize the Review monitor display to the acquiring study 3 Open previous study 4 Close the study 5 Display study information 6 Display comment menu Rel...

Страница 37: ...nter Perform calibration according to the following procedure if needed To perform calibration display the image of reference object such as a catheter previously acquired with the same magni cation a...

Страница 38: ...est part of the stenosed artery The stenosis is measured and indicated on screen 4 Specify the reference value 1 Click button from Manual menu 2 Click points on opposite sides of the part of the un st...

Страница 39: ...measured object is located on the ISO center Perform calibration according to the following procedure if needed To perform calibration display the image of reference object such as a ball previously a...

Страница 40: ...rve passing through clicked points is created Move a point by clicking and dragging it to a new position Delete a point by clicking it with the left and right mouse buttons simultaneously 3 Once the o...

Страница 41: ...n 3 Create peak hold image 4 Set ROI of MAP 5 Resetting mask image 6 Executing reregistration Shift the mask image 7 Create a contour enhanced blood vessel image from DSA image No Button Function 1 2...

Страница 42: ...42 1 Click and select Printer dialog will appear 2 Setting any required items such as 3 Add images with 4 Click Printing is started 1 Click from Study menu Printing Images Closing Study...

Страница 43: ...the search target set by the System and Drives buttons plus all search criteria 2 View Displays the selected study or studies in the Image Viewer window 3 RDSR Available if there is dose report Displa...

Страница 44: ...he con rmation window appears click OK Sending Studies No Button Function 6 Info Displays detailed information about the selected patient 7 Delete Permanently deletes the unprotected studies selected...

Страница 45: ...writer 3 Insert a blank disk Media Writing dialog box appears Volume name is changeable Select the writing item from Images RDSR or Both 2 Click Writing is started To cancel a media write click on the...

Страница 46: ...information display to ensure normal operation NOTE Safety Switch Touch sensors are mounted on the FPD front cover and the collimator cover When these sensors are tripped the movement of C arm and cat...

Страница 47: ...procedure Digital Operation Cabinet Control Room 1 Press and hold the Power button for at least 5 seconds The power goes off Cabinet Equipment Room 2 Open the switch cover in front of the cabinet and...

Страница 48: ...on the power After the application screen is displayed on both acquisition and review monitors shut down the system again by standard procedure and then restart the system NOTE Digital Angiography sys...

Страница 49: ...ll Modes in the Calibration Type list 4 Click Progress messages are displayed during calibration Completion message is displayed when calibration is completed 5 Click Click if you terminate the calibr...

Страница 50: ...er failure occurs or the longer one once shut down the system and then re activate the system according to the Startup procedure P 16 or Chapter 3 on DAR 9500f Operation Manual M517 E181 The system is...

Страница 51: ...An error occurred during cache space reservation If CD R or DVD R media are used as patient data archive archive all unsaved patient data to new CD R or DVD R No available cache Cannot retrieve media...

Страница 52: ...system monitor errors in the following list are also displayed Stop study immediately and contact Shimadzu service representative Action Message Description Sensor Temperature error Cooling Unit Wate...

Страница 53: ...the focus is switched to normal one if this message appears after rebooting the X ray high voltage generator contact Shimadzu service representative No Filament S L Both Abnormal lament If this messa...

Страница 54: ...aphy is not possible while this message is displayed Action Message Description HU Over Predicted Value predicted to go out of range Continuing radiography with present setting will cause heat unit ov...

Страница 55: ...ing displays Starter is not working The starter is in standby status Contact Shimadzu service representative Error Messages at Power ON Action Message Description Contact Shimadzu service representati...

Страница 56: ...ntact Shimadzu service representative Motor Error XXX Motor command fail to drive the motor Camera Rotation Error Error of FPD rotation potentiometer Collimator Error XXX Error of the collimator poten...

Страница 57: ...Error Break in the stop switch circuit is detected One of the doubled circuits is out of order Although the equipment is available for the temporary use contact Shimadzu service representative immedi...

Страница 58: ...of next Fluoro acquisition Rad Rad Program Associated prede ned Rad mode selected upon selection Toggle Rad Mode Biplane only Change the plane in order of Frontal Bi and Lateral Set Rad Mode BiPlane B...

Страница 59: ...Zoom level decreases slightly upon selection Flip image Horizontal Flip an acquisition image horizontally Flip image Vertical Flip an acquisition image vertically Play Play Sync Play ON Play Single Lo...

Страница 60: ...utton Function A R Annotations Show Annotation Properties Annotations properties shown upon selection Pointer Pointer tool selected upon selection A relevant to Acquisition monitor R relevant to Refer...

Страница 61: ...http www shimadzu com...

Страница 62: ...Operation Guide BRANSIST alexa Digital Angiography System...

Отзывы: