11. Environment Settings
11.3 Stability Detection Band
The appearance of the Stability Mark (
) indicates that the mass display has been stabilized. The
condition for judging stability is user-selectable. When the stability detection band is set to 1, the stabil-
ity mark appears when the mass display stays within 1 count for a set period of time. The stability
detection band setting may be selected from three levels: 1 count, 5 counts and 10 counts. The default
setting is 1 count. One count is equal to 0.1mg when the minimum display is 0.1mg.
The setting of the stability detection band can be checked in the settings check display (see
pressing the
[CAL]
key four times from mass display.
Setting the Stability Detection Band
1
Press the
[CAL]
key repeatedly from the mass
display. When “FUnC.SEL” appears, press the
[O/T]
key. “CAL” is displayed.
2
Press the
[CAL]
key twice. “bAnd:
∗∗
” is shown.
The
∗∗
part (single or double digits) shows the
current stability detection band setting.
3
Press the
[O/T]
key. “b-1” appears. At this
point, pressing the
[CAL]
key changes the display
in this order: “b-1”, “b-5”, “b-10” representing sta-
bility detection bands of 1 count, 5 counts, and 10
counts, respectively. When the current setting is
displayed, the stability mark (
) appears.
4
Press the
[O/T]
key to select the stability detec-
tion band currently displayed, or
5
Press the
[POWER]
key to return to the
“bAnd:
∗∗
” display without changing the setting.
6
Press the
[POWER]
key to return to the mass
display.