Shimadzu AOC-20i Скачать руководство пользователя страница 16

Shimadzu AOC-20

5 Sequence

5 Sequence

To use the autosampler in the

Clarity

sequence, following conditions

must be met:

The sequence must be saved
The sequence must be set as

Active Sequence

All used methods must have the

External Start

enabled and the

Autostop

time set

Method Setup - AS

dialog in the used methods must be filled

The required values for starting vial (

SV

), ending vial (

EV

), number of

injections from vial (

I/V

) and injected volume are set in the sequence

table.

Standard sequence operation

After sequence start, the AS control method will be sent to the sampler
and the datastation will wait for the injection signal from sampler. After
receiving it, the run starts and after finishing, the method for next injection
will be sent to the sampler.

Note:

Autosampler control is not available during Single Run operation.

- 12 -

Содержание AOC-20i

Страница 1: ...AOC 20 Clarity Control Module ENG Code Rev M090 50E Date 24 10 2017 Phone 420251013400 DataApex Ltd Fax 420251013401 Petrzilkova 2583 13 clarity dataapex com 15800Prague 5 www dataapex com The Czech...

Страница 2: ...emarks of DataApex Ltd Microsoft and WindowsTM are trademarks of Microsoft Corporation DataApex reserves the right to make changes to manuals without prior notice Updated manuals can be downloaded fro...

Страница 3: ...Requirements 2 3 Installation Procedure 3 3 1 Autosampler Setup communication 3 3 1 1 Connections 4 3 2 Clarity Configuration 5 4 Using the control module 7 4 1 Hardware Configuration dialog 7 4 2 Met...

Страница 4: ...RK1 capitals indicates the name of the file and or directory ACTIVE capitalitalics marks the state of the station or its part The bold text is sometimes also used for important parts of the text and t...

Страница 5: ...autosamplers The control module enables direct control of the instrument over serial line Fig 1 AOC 20i Direct control means that the autosampler can be completely controlled from the Clarity environm...

Страница 6: ...MultiCOM adapter p n MC01 is available Serial cross DB9F DB9F cable p n SK01 Note Cables are not part of Clarity Control Module It is strongly recommended to order required cables together with the Co...

Страница 7: ...he RS232 compliant serial cross DB9F DB9F cable Fig 2 Serial cross DB9F DB9F cable The RS 232 cable is plugged in to the autosampler controller board not the tower present at the back side of the GC r...

Страница 8: ...The Clarity Ready OUT digital output is not used the commands are transmitted over the serial line Refer to Shimadzu manual for description of cables supplied with the autosampler When autosampler ha...

Страница 9: ...con on the desktop Invoke the System Configuration dialog accessible from the Clarity window using the System Configuration command Press the Add button see Fig 4 on pg 5 to invoke the Available Contr...

Страница 10: ...atus of the autosampler will appear in the Connection row Note The Shimadzu AOC 20i s Setup dialog is more closely described in the chapter Shimadzu AOC 20i s Setup on pg 10 The AOC 20 autosampler ite...

Страница 11: ...r transfer of methods from and to autosampler Caution If you have injection methods already established in the sampler it is advisable to download them to Clarity using the From AS button and save the...

Страница 12: ...before injection Valid range 0 99 Rinse with solvent Solvent washing times of the syringe after injection Valid range 0 99 Viscosity time Waiting time of the plunger after sample suction Valid range 0...

Страница 13: ...speed Injection speed of the syringe Select Solvent Selects the appropriate solvent for solvent wash in case there are three solvent vials on the sampler rack It is necessary to enable this option in...

Страница 14: ...ler use the Function key to check modify them Default values Baud rate 2400 F80 2 Parity none F81 0 Stop bits 2 F82 1 Note The changes made on the sampler are active only after power restart Connect B...

Страница 15: ...s the type of the syringe AOC 17 mode Setting this checkbox enables AOC 17 compatible mode To use the AOC20 in this mode set the functions F84 1 default is F84 0 Note Regardless of mode used the Chann...

Страница 16: ...s must be filled The required values for starting vial SV ending vial EV number of injections from vial I V and injected volume are set in the sequence table Standard sequence operation After sequence...

Страница 17: ...ing lines should be added COM1 echo on textmode on filename CommDrv_ D txt reset off Note Instead of COM1 type the correct serial port used to communicate with the Shimadzu AOC 20 detector Note D or d...

Отзывы: