SA5041 Digital Multimeter User ’s Guide
Shijiazhuang Suin Instruments Co., Ltd 18
ACI
, you must send the command with parameter 1or ON as below,
[:SENSe]:CURRent:AC:RANGe:AUTO 1
Parameter types:
<b> Boolean: Used to enable or disable an instrument operation. 0 or OFF
disables the operation; 1 or ON enables the operation.
CURRent:AC:RANGe:AUTO 1 enable AUTO range
<name> Name parameter: Select a parameter name from a listed group.For example,
:FUNCtion <name>
<name> = “VOLTage:AC” select AC voltage
“VOLTage[:DC]” select DC voltage
“CURRent:AC” select AC current
“CURRent[:DC]” select DC current
“RESistance” select 2-line resistance
“FREQuency” select frequency
“DIODe” select diode testing
“CONTinuity” select continuity testing
“TEMPerature” select temperature testing
<n> Numeric value
:
This parameter stands for an integer (take 6 for example),
numbers of real number (take 25.3 for example)
[SENSe[1]:CURRent[:DC]:NPLCycles 1
[SENSe[1]:CURRent[:DC]:NPLCycles 0.1
Command abbreviation rules
1. If the command length less than or equal to four characters, there is no abbreviation.
For example,
:AUTO=:AUTO
2. The rule is available for commands which longer than four characters.
3. If the fourth characters is v,o,w,e,l, then cancel it and all characters behind them.
For example,
:IMMediate=:IMM
4. Special rules – the abbreviation of the below command only uses the first two