CHAPTER 3. MAINTENANCE
If you put the lamp in the opposite direction,
it will apparently work, but the lifetime will
be significantly shorter (it will become darker
and darker over the time).
Put back the cover, taking care of the servo cable:
Tighten the three screws - that’s it.
3.2
Replacing the flat lamp
Before any maintenance on the calibration
module, switch off any power supply (re-
move the plug).
Remove the screw in the middle of the cover, and
loosen the other two:
Remove the cover:
Unplug the servo cable:
Remove four screws around the module:
Gently slide the electronics from the calibration mod-
ule body. Take care to protect the halogen lamp dur-
ing this operation: