13
|
P a g e
INSTALLATION
D
EIONIZED AND
D
ISTILLED
W
ATER
Do not use deionized water
to clean or humidify the incubator. Use of deionized water may corrode
metal surfaces and voids the warranty. The manufacturer recommends the use of distilled water in
the resistance range of 50K Ohm/cm to 1M Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0
uS/cm, for cleaning and humidifying applications.
I
NSTALL THE
S
HELVING
To install the incubator shelves perform the following:
1.
Hold the shelf slide at about a 10° angle. See figure 2A
2.
Insert the hooked tab into the vertical slot at the desired height (Figure 2B)).
3.
Push the slide into the horizontal slots.
4.
Repeat for shelf slide at the same height on the opposite side of the incubator.
5.
Place a shelf on the shelf slides. Repeat this procedure for additional shelves as needed.
Shelving Installation
A
CCESS
P
ORT
SMI incubators are provided with a plastic cap for the access port on the right side of the
incubation chamber. Always leave this cap in place, except when introducing sensor
probes into the chamber, such as those used for temperature calibrations. Removing the
cap during normal operations can adversely impact temperature stability and uniformity.
Содержание SMI11
Страница 1: ...SMI INCUBATORS 100 120 Voltage Installation Operation Manual SMI2 SMI6 SMI7 SMI11 SMI12...
Страница 2: ...2 P a g e Pictured on cover SMI6 SMI2 SMI7 SMI11 SMI12 2 Stacked SMI6s...
Страница 14: ...14 P a g e INSTALLATION...
Страница 16: ...16 P a g e GRAPHIC SYMBOLS...
Страница 18: ...18 P a g e CONTROL PANEL OVERVIEW...
Страница 32: ...32 P a g e USER MAINTENANCE...