The SDHmini will display a moving bar until it assesses that
the sensor has stabilised (up to 10 minutes).
The full scale value, -10 °C dewpoint, is now displayed.
Press when complete.
Close the desiccant head when instructed to do so and
press .
Pressing during
CalSet
aborts the
process.
The value present at the start of the
aborted
CalSet
process is reinserted.
The SDHmini will return to the normal operational mode.
6.4 Calibration Notes
If the instrument fails calibration, a message will be displayed on the screen
until acknowledged by the user, e.g.
AutoCal Failed
Sensor Open Circuit
or
AutoCal Failed
Sensor Short Circuit
This will mean replacing the sensor or its interconnecting cable in some cases.
The entered
AutoCal
value must not be pressure compensated.