- Mechanical installation

    2.3.4 Reference pipe installation

  Disturbing factors that may influence the level measurement in liquids, as for example:

    -  foam presence on the product surface (Fig.12)

    -  internal structures presence in the tank (Fig.13)

    -  presence on the liquid surface of floating bodies (Fig.14)

  can be avoided with the use of level measurement inside of pipes (by-pass pipe or calm pipe with 100mm min. 
 

diameter for PTU50-51, or 125mm min. diameter for PTU56

)

  The pipe must have a length greater or equal than the empty distance, also, must have some of vent holes 

  (Fig. 14-A) to allow the pipe regular filling and emptying.  

 

In the programming menu, to the “

PRODUCT

” parameter, must select the “

LIQUID PIPE

” option (see page 7 or 11)

  2.3.5 Agitators presence

  The level measurement is possible thanks to the  

Auto-Tuned

 statistical filter.

  Should rarely need to adjust the filter setting by editing 2 

PTU50-51-56 

sensor    

programming parameters:

   - 

FILTER

; this parameter is present in the 

Quick Setup 

menu (page 8)  

 

    and in the Advanced Configuration “

SETUP

” menu (page 11); 

      increasing the parameter value, decreases the sensor sensitivity to the  

 

    level measurement sudden variations. 

   - 

F-WINDOW

; this parameter is present in the Advanced Configuration  

 

   “

SERVICE

” menu (page 18); decreasing the parameter programmed    

 

  value, increases the sensor immunity to false echoes.

Fig.12

Fig.13

Fig.14

Fig.14-A

Fig.15

  2.3.6 Mechanical installation accessories

Fig.16

Содержание PTU5 Series

Страница 1: ...Certificate In conformity to the company and check procedures I certify that the equipment PTU Production and check date Serial n is conform to the technical requirements on Technical Data and it is...

Страница 2: ...PTU51 and PTU56 units are compact sensors and have a via connector quick connection The IP68 protection makes them suitable for external applications with direct exposure to the weather including are...

Страница 3: ...for 12m max PTU56 range where the sensor can not measure PTU50 1 5m PTU51 6m PTU56 12m D mm Lmax mm D mm Lmax mm 100 80m 125 240 125 240 150 300 150 300 Fig 10 Fig 11 Tab 2 2 3 Mounting precautions 2...

Страница 4: ...illing and emptying In the programming menu to the PRODUCT parameter must select the LIQUID PIPE option see page 7 or 11 2 3 5 Agitators presence The level measurement is possible thanks to the Auto T...

Страница 5: ...tighten the connector ring nut position the cable so that it forms a downward curve at the M20 output Fig 18 in this way the condensation and or rain water will tend to drip from the curve bottom Fig...

Страница 6: ...ification OK BK Exit configuration Back to previous menu Eco map from RUN mode Parameters values selection Parameters scroll Fig 21 PTU5x Configuration and Quick Start 5 QUICK START MODE From RUN mode...

Страница 7: ...U50 1 5m max 300mm PTU51 6m max 500mm PTU56 8m max 5 1 Struttura men di configurazione rapida 4mA distance 20mA distance 100 0 Level measurement PTU50 51 56 referent point Fig 22 5 1 1 SET DISTANCE 4m...

Страница 8: ...II 49 20090 Rodano MI ITALY tel 39 0295328257 fax 39 0295328321 web www sgm lektra com e mail info sgm lektra com Pagesfrom9to20 fullmanual canbedownloadedfromourwebsitewww sgm lektra comintheproducts...

Отзывы: