4.0
Function Control and Adjustment
According to German Safety Standards (DVGW-Arbeitsblatt
G 465/IV) tests, controls and inspections of gas detectors are
obligatory.
Paragraph 5 of these standards „Control and tests of Equipment“
requests that „all gas measuring equipment used for monitoring
has to be checked before an operation.“ This check includes the
test of zero and gas indication.
According to paragraph 6.1 „Inspections“ these have to be carried
out „depending on the frequency of operations ... at least ... once a
year.“
These inspections include zero control and a sensitivity test.
The results have to be recorded.
The following equipment is required for the function control:
- the charged
EX-TEC® SNOOPER L
,
- the test set type
SPE
SPE
SPE
SPE
SPE
(Pict. A),
- the adapter head
- a test gas (1.00 Vol.% of CH
4
).
Pict. A
- Test Set Type
SPE
SPE
SPE
SPE
SPE
B e t r i e b s a n l e i t u n g d e r
G e r ä t e b e a c h t e n !
Z u r Ü b e r p r ü f u n g v o n
G a s m e ß g e r ä t e n m i t P u m p e
g e m ä ß D V G W - R i c h t l i n i e
G 4 6 5 / I V
3
1
2
Содержание EX-TEC SNOOPER L
Страница 19: ...Vers 02 29 04 1994 21 Pr fungsscheinPTB Nr Ex 94 Y 4601 Anhang A...
Страница 20: ...22 Vers 02 29 04 1994 Pr fungsscheinPTB Nr Ex 94 Y 4601 Anhang A...
Страница 22: ...24 Vers 02 29 04 1994 NOTIZEN...
Страница 31: ...42 Vers 02 29 04 1994 Test CertificatePTB No Ex 94 Y 4601 Annexe A...
Страница 32: ...44 Vers 02 29 04 1994 NOTES...