48
7 Adjustment menu
7.6
Setting the CH
4
sensor
CH
4
- zero point 0 %VOL
now release
fresh air
from the
SPE VOL
wait for the display to settle at
a stable value and confirm the
adjustment with the on/off key
(
OK
appears in the LCD)
pressing the
arrow-up key
brings you to the next display
CH
4
- sensitivity 2.20 %VOL
now release
test gas 2.20 %VOL
(50 %UEG) CH4
from the SPE
VOL (cf. section 7.8: Setting
the test gas concentration)
wait for the display to settle at
a stable value and confirm the
adjustment with the on/off key
(
OK
appears in the LCD)
once this has happened, turn
off the test gas feed
pressing the
arrow-up key
brings you to the next display