![seward Stomacher 80 Biomaster Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/seward/stomacher-80-biomaster/stomacher-80-biomaster_user-manual_1250359006.webp)
5. WARRANTY AND SERVICE POLICY
All Seward Stomachers® are protected by a 36-month warranty covering faulty
workmanship and materials. In the event of a fault developing within this period
please contact Seward Customer Service who will organise the collection/return
of the instrument for warranty repair and/or replacement at Seward’s discretion.
The warranty is void if the instrument has been damaged as a result of
unauthorised service, modification or subject to neglect, misuse, and accident,
improper installation or operation.
6. COMPLIANCE
Seward Stomachers® are designed and manufactured to conform to the
following standards: -
• IEC 61010-1 : 1990 Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use.
• This equipment complies with the essential requirements of the European
Council Directive 2006/42/CE relating to EMC and carries the CE mark.
• A full list of compliance is detailed on our declaration of conformity and
available on our website or from our customer services team.
7. STORAGE AND TRANSPORT
Recommended storage and transport conditions are:-
Ambient Temperature Range -20 to + 50°C
Relative Humidity Range 10 - 90%
Seward Stomachers® should ideally be stored in a clean environment and for
long periods of storage it should be re-packed in the original packaging.
Seward Stomachers® should be transported in the original packaging.
8. STOMACHER® 80 BIOMASTER - SPECIFICATIONS
Model
American and
Japanese Product
EU Product
Cat No.
0080/000/AJ
0080/000/EU
Dimensions
290mm W x 325mm D x
190mm H (unpacked)
290mm W x 325mm D x
190mm H (unpacked)
Weight
15kg (unpacked)
15kg (unpacked)
Supply Voltage
100-120V
220-250V
Supply Frequency
50-60Hz
50-60Hz
Mains Inlet Fuses*
T2.0A 250V 5x20mm, break-
ing capacity: 100A @ 250V
T800mA 250V 5x20mm,
breaking capacity: 35A @
250V
Transformer Fuse*
T250mA 250V 5x20mm,
breaking capacity: 35A @
250V
T250mA 250V 5x20mm,
breaking capacity: 35A @
250V
PCB Fuse*
F2.5A 250V 5x20mm, break-
ing capacity: 100A @ 250V
F2.5A 250V 5x20mm, break-
ing capacity: 100A @ 250V
Insulation
Class 1 (Earthed)
Class 1 (Earthed)
Installation
Over Voltage Category II
Over Voltage Category II
Pollution
Pollution Degree 2
Pollution Degree 2
Duty Cycle
Continuous
Continuous
Rated Load
55W
55W
Disposable Bag
152mm x 101mm
152mm x 101mm
Minimum Recommended
Capacity
5ml
5ml
Maximum Recommended
Capacity
80ml
80m
Maximum Recommended
Sample Temperature
60°C
60°C
Recommended Operating
Ambient Temperatur Range
10-35°C
10-35°C
Recommended Operating
Relative Humidity Range
10-89%
10-89%
Fixed Timer Settings
30 seconds (±5%)
60 seconds (±5%)
120 seconds (±5%)
Continuous
0 seconds (±5%)
60 seconds (±5%)
120 seconds (±5%)
Continuous
No Load Paddle Speeds
(@ nominal 20°C)
230rpm (±5%) LOW
265rpm (±5%) NORM.
300rpm (±5%) HIGH
230rpm (±5%) LOW
265rpm (±5%) NORM.
300rpm (±5%) HIGH
Paddle Clearance
0-l0mm
0-l0mm
* Only these or equivalent UL-approved fuses should be used
10
9