seward Stomacher 80 Biomaster Скачать руководство пользователя страница 6

5. WARRANTY AND SERVICE POLICY

All Seward Stomachers® are protected by a 36-month warranty covering faulty 
workmanship and materials. In the event of a fault developing within this period 
please contact Seward Customer Service who will organise the collection/return 
of the instrument for warranty repair and/or replacement at Seward’s discretion.

The warranty is void if the instrument has been damaged as a result of 

unauthorised service, modification or subject to neglect, misuse, and accident, 

improper installation or operation.

6. COMPLIANCE

Seward Stomachers® are designed and manufactured to conform to the 
following standards: -

•     IEC 61010-1 : 1990 Safety Requirements for Electrical Equipment for  

 

       Measurement, Control and Laboratory Use.

•     This equipment complies with the essential requirements of the European  
       Council Directive 2006/42/CE relating to EMC and carries the CE mark. 

•  A full list of compliance is detailed on our declaration of conformity and 

available on our website or from our customer services team.

7. STORAGE AND TRANSPORT

Recommended storage and transport conditions are:-

Ambient Temperature Range -20 to + 50°C

Relative Humidity Range 10 - 90%

Seward Stomachers® should ideally be stored in a clean environment and for 
long periods of storage it should be re-packed in the original packaging.

Seward Stomachers® should be transported in the original packaging.

8. STOMACHER® 80 BIOMASTER - SPECIFICATIONS

Model

American and 

Japanese Product

EU Product

Cat No.

0080/000/AJ

0080/000/EU

Dimensions

290mm W x 325mm D x 
190mm H (unpacked)

290mm W x 325mm D x 
190mm H (unpacked)

Weight

15kg (unpacked)

15kg (unpacked)

Supply Voltage

100-120V

220-250V

Supply Frequency

50-60Hz

50-60Hz

Mains Inlet Fuses*

T2.0A 250V 5x20mm, break-
ing capacity: 100A @ 250V

T800mA 250V 5x20mm, 
breaking capacity: 35A @ 
250V

Transformer Fuse* 

T250mA 250V 5x20mm, 
breaking capacity: 35A @ 
250V

T250mA 250V 5x20mm, 
breaking capacity: 35A @ 
250V

PCB Fuse*

F2.5A 250V 5x20mm, break-
ing capacity: 100A @ 250V

F2.5A 250V 5x20mm, break-
ing capacity: 100A @ 250V

Insulation

Class 1 (Earthed)

Class 1 (Earthed)

Installation

Over Voltage Category II

Over Voltage Category II

Pollution

Pollution Degree 2

Pollution Degree 2

Duty Cycle

Continuous

Continuous

Rated Load

55W

55W

Disposable Bag

152mm x 101mm

152mm x 101mm

Minimum Recommended 
Capacity

5ml

5ml

Maximum Recommended 
Capacity

80ml

80m

Maximum Recommended 
Sample Temperature

60°C

60°C

Recommended Operating
Ambient Temperatur Range

10-35°C

10-35°C

Recommended Operating 
Relative Humidity Range

10-89%

10-89%

Fixed Timer Settings

30 seconds (±5%)
60 seconds (±5%)
120 seconds (±5%)
Continuous

0 seconds (±5%)
60 seconds (±5%)
120 seconds (±5%)
Continuous

No Load Paddle Speeds
(@ nominal 20°C)

230rpm (±5%) LOW
265rpm (±5%) NORM.
300rpm (±5%) HIGH

230rpm (±5%) LOW
265rpm (±5%) NORM.
300rpm (±5%) HIGH

Paddle Clearance

0-l0mm

0-l0mm

* Only these or equivalent UL-approved fuses should be used

10

9

Отзывы: