Chapter 5:
Timing
98
v3.3-a 12/07/2022
timing source. Then the user can allow to trigger the
holdover between W
3.7210.1 Time to expire
<Integer>
Default: 79800
Time until the holdover is considered out of
specification and expired (default ~24h)
3.7210.2 Force Trigger
<Enum>
0.
STOP
1.
START
2.
NONE
Force to manually trigger the holdover (START) or to
expire it (STOP) without waiting the expiration timer.
NONE, does nothing.
HO in between timing source
Using the source rank the HO can be placed between two timing sources. For example,
the user can use WR as primary timing source, HO as secondary and PTP 8265.1 as third
one. This means that if WR fails, the device will enter in HO until expiration and finally
switch to PTP that might provide better accuracy than an expired HO.
5.8.2 Info/Overview
An overview of the holdover timing source is provided to monitor its state at any time as shown
in Figure .
If its state is UNAVAILABLE this means that the holdover oscillator has not been detected and its
related information in irrelevant. If you have ordered the holdover option but the device does
not detect it, please contact [email protected].
Figure 5:23 - Holdover overview meanwhile learning
OID
Name
Value Type
Description
3.7220.x holdover/info/xxx
Information about the Holdover timing source
3.7220.2 State
<Enum>
0.
UNAVAILABLE
1.
DISABLED
2.
LOCKING
3.
LEARNING
4.
READY
Current state of the holdover timing source as
explained in the introduction of the section. If
the holdover is not detected the corresponding
state if UNAVAILABLE. The user can also
manually force it as DISABLED in case to avoid
triggered it.
Содержание WR-Z16
Страница 2: ...Chapter 1 Introduction 2 v3 3 a 12 07 2022 ...
Страница 10: ...Chapter 1 Introduction 10 v3 3 a 12 07 2022 ...
Страница 72: ...Chapter 5 Timing 72 v3 3 a 12 07 2022 ...
Страница 148: ...Chapter 8 Device Maintenance 148 v3 3 a 12 07 2022 ...
Страница 159: ...Chapter 9 Troubleshooting 159 v3 3 a 12 07 2022 Severity info local0 rsyslog server1 IP local0 rsyslog server2 IP ...