background image

7

4.0 Operation

For most accurate pressure measurement, allow the Model 730 to warm up for at least 15 
minutes. After installation, periodically check the zero output reading to verify correct output. 
Adjust the zero potentiometer if incorrect (See Section 5 for zero adjustment instructions). 
The signal output of the Model 730 is linear with pressure; e.g., for a 10 VDC FS Model 730, 10 
VDC equals 100% FS output; 1 VDC equals 10% FS output.

Table 1 indicates the lowest pressures recommended for reading and pressure control for 
each range of the Model 730. The lowest suggested pressure for reading is limited by the 
resolution and the accuracy of the Model 730. This is directly related to the electrical noise on 
the signal output and can be significantly effected by incorrect electrical ground connection, 
or connection to an electronically noisy power supply or readout instrument. Improved results 
may be obtained if the transducer is operated in an environment with stable temperature and 
air flow. The lowest recommended pressure used for control applications, such as a closed 
loop downstream pressure control system, is based on a minimal signal output of 50 mV. 

NOTE: 

if the unit has been exposed to sudden environmental changes, allow at least 2 hours 

before making any adjustments.

Table 1: Recommended Lowest Pressures

Available for reading and pressure control

Full scale range

Lowest recommended 
pressure reading

Recommended lowest 
pressure for control

10 Torr

0.005 Torr

0.05 Torr

20 Torr

0.010 Torr

0.10 Torr

100 Torr

0.050 Torr

0.50 Torr

200 Torr

0.100 Torr

1.00 Torr

1000 Torr

0.500 Torr

5.00 Torr

10 mbar

0.005 mbar

0.05 mbar

20 mbar

0.010 mbar

0.10 mbar

100 mbar

0.050 mbar

0.50 mbar

1000 mbar

0.500 mbar

5.00 mbar

1 kPa

0.0005 kPa

0.005 kPa

2 kPa

0.0010 kPa

0.010 kPa

10 kPa

0.0050 kPa

0.050 kPa

100 kPa

0.0500 kPa

0.50 kPa

Содержание Model 730

Страница 1: ...Model 730 Operating Instructions Setra Systems Inc 159 Swanson Road Boxborough MA 01719 800 257 3872 www setra com...

Страница 2: ...ble capacitor within a small reference vacuum chamber As the pressure increases the diaphragm deflects and the gap between the electrode and diaphragm reduces causing an increase in the capacitance Th...

Страница 3: ...to connect the NW16 flange to an NW10 flange on the system Use a Male 4 VCR or 8 VCR Face Seal Fitting and sealing gasket to connect to the female swivel 4 VCR or 8 VCR Style Face Seal Fitting NOTE Ti...

Страница 4: ...with 5 pin terminal strip and 0 5 and 0 25 OD Tube 0 99 25 0 50 13 3 05 77 3 47 87 0 75 19 0 25 6 3 17 80 1 50 38 Pin 1 Mating connector 0 12 3 05 Pin 1 Terminal strip Model 730 with 0 5 diameter tube...

Страница 5: ...0 13 3 24 82 0 75 19 0 25 6 3 0 76 Model 730 with 0 5 diameter tube Model 730 with 0 25 diameter tube Figure 3 Outline drawing of Model 730 with 6 cable and 15 pin D sub and 0 5 and 0 25 OD Tube Pin 1...

Страница 6: ...a metal case and it must be in direct contact with the cable braided shield for the entire circumference of the cable A single wire from the shield to the connector case will not be effective 3 The c...

Страница 7: ...o an electronically noisy power supply or readout instrument Improved results may be obtained if the transducer is operated in an environment with stable temperature and air flow The lowest recommende...

Страница 8: ...model 730 to be 0 001 to 0 001 V Make this adjustment at a pressure at least 1 2 decade below the Model 730 s resolution e g for a 10 Torr FS unit the zero pressure should be less than 5E 5 Torr For...

Страница 9: ...e Solution No signal output Incorrect or no supply voltage Ensure power supply is used as specified in Section 3 Readout display short circuit or incorrect impedance Ensure impedance of readout unit i...

Страница 10: ...Storage temperature 20 to 85 C Electrical data voltage Excitation Output4 12 to 30 VDC for 0 10 VDC 9 to 30 VDC for 0 5 VDC Current consumption 10 mA max Output load 10 k Load Output impedance 1 ohm...

Страница 11: ...ective in materials or workmanship within the warranty period provided that a the product has not been subjected to abuse neglect accident incorrect wiring not our own improper installation or servici...

Страница 12: ...Setra Systems Inc 159 Swanson Road Boxborough MA 01719 800 257 3872 www setra com SS730 REV C 1 2019...