background image

 

 

 

 

OPERATING INSTRUCTIONS :

 

INSTALLATION :

 

When unpacking the instrument, inspect for any damage during transit. Any such damage 

should immediately be intimated to the factory for appropriate action. 

The  instrument  is  light  weight  and  portable  and  does  not  require  any  permanent 

installation. Before connecting to the mains supply, it should be ensured that the mains voltage is 

between 200 to 240 volts a.c. and a 

good earth connection is available in the power socket.

 

 

OPERATING INSTRUCTIONS : 

1. 

Ensure  that  the  Main's  switch  is  in  'OFF'  position.  Connect  the  instrument  to  a  three  pin 

well earthed

 main's point. 

2. 

Switch  'ON'  the  instrument.  Allow  one  minute  for  warm  up,  stabilisation  of  components 

and circuits. 

3. 

Using 'Zero adj' adjust the panel meter reading to zero. 

4. 

Connect the coaxial cable to the source and select appropriate range. 

 

PRECAUTIONS :

 

1. 

The  instrument  should  be  placed  as  near  to  the  source  as  possible.  This  will  minimise 

current leakage paths, noise pick-ups and capacitive loading. 

2. 

Use  only  the  best  quality  coaxial  cable  with  amphenol  connectors  at  both  ends.  Cable 

system should be made as rigid and vibration free as possible, since cable movement can 

cause various noise signals, which are significant in the high impedance systems. 

3. 

Many instrumentation situations, such as flame detectors in gas chromatographs, involve 

measurement of low level currents from high voltage sources. If the source is not current 

limited  it  requires  input  protection.  All  that  is  required  is  a  resistor  in  series  with  the 

affected  input  terminal  so  that  the  maximum  overload  current  is  0.5  mA  (for  example 

200  K  ohms  for  a  100  volt  overload).  The  instrument  is  protected  for  200  volts  and  a 

suitable series resistor should be added in the source circuit for additional voltage of the 

source. This will not cause any significant reduction in the performance. 

4. 

Please note that the instrument is essentially designed for very low current measurement 

from high impedance source. Measurement of currents from comparative low impedance 

source  may  require  zero  adjustment  even  at  higher  ranges.  This  may  be  checked  and 

corrected if necessary, for a particular load. 

Отзывы: