![Servomex xendos 1800 Series Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/servomex/xendos-1800-series/xendos-1800-series_quick-start-manual_1247054013.webp)
xendos 1800 Series
6
3
Do not exceed the maximum current output load impedance of 600
Ohms.
4
Electrical power should be supplied at 100V to 240Vac ± 10% - 50/60Hz.
(Note: The optional internal sample pumps are AC voltage dependant.
Ensure that the correct AC input voltage is connected to analysers fitted
with a sample pump. A power on LED is illuminated when power is
present at the analyser).
5
For compliance with the EMC standards, the external earth / ground
connection must always be connected to a local earth / ground. Refer to
Figure (A). In addition, the connections to TB2, TB3 and TB4 must
always be made using a screened or shielded cable with the earth
termination being made at the cable gland. Ensure good EMC practices
are observed when making connections.
6
All electrical connections and access to the power supply fuse, F1, are
made to the terminal board inside the right hand side of the analyser.
Refer to Figure (B) for details and full specifications are given in the
following table. To gain access refer to Figure (A) and remove the right
hand cover by unbolting the four M6 bolts using a 5mm allen key. After
electrical connections are complete ensure the lid is fully bolted down.
WARNING
If the relay connections to TB2 are used to switch a combination of
240V ac and 28V dc signals, then all parts of the dc relay circuits
must be protected to a minimum of IP20
Содержание xendos 1800 Series