66
© Servomex Group Limited. 2020
01910001B_8
Alarm mode
Effect of hysteresis
Low alarm
Once the low alarm condi�on has been ac�vated, the alarm condi�on will not
be deac�vated un�l a sample measurement is above (alarm level + hysteresis
level).
High alarm
Once the high alarm condi�on has been ac�vated, the alarm condi�on will not
be deac�vated un�l a sample measurement is below (alarm level
-
hysteresis
level).
For example:
•
If a ‘low’ alarm has an alarm level of 18% and a hysteresis level of 1%, the alarm will be
ac�vated when a sample measurement is < 18%, and the alarm will not be deac�vated
un�l a sample measurement is >
19%.
•
If a ‘high’ alarm has an alarm level of 20% and a hysteresis level of 2%, the alarm will be
ac�vated when a sample measurement is > 20%, and the alarm will not be deac�vated
un�l a sample measurement is < 18%.
5.20.4
Viewing the measurement alarm
status
1.
Select
View.
2.
In the Alarm status screen shown in Figure
both measurement alarms are shown as
"Inac�ve"; that is, either the mode of each alarm is set to ‘none’, or no alarm condi�on
currently exists.
Figure
66 -
The Alarm status screen
If a measurement alarm condi�on exists when you view this screen, the screen will show:
•
The alarm number ("1" or
"2").
•
The sample reading which triggered the alarm
condi�on.
•
The alarm mode (where "<" indicates a low alarm and ">" indicates a high alarm).
•
The alarm
level.
5.20.5
Setting the Measurement Filter
Level
Use the following procedure to set up the measurement filtering:
1.
Select:
Set Up
Filtering
2.
The current filter level is shown.
Содержание SERVOTOUGH Oxy
Страница 2: ...This page inten onally blank...
Страница 4: ...This page inten onally blank...
Страница 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Страница 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Страница 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...