30 DF-760E
Installation, Start Up and Shut Down
6.
The state of the gas control valves for each of the sensors is indicated on the main
display. The default state of a factory-shipped instrument is isolation for both
sensors. This can be adjusted in the Power Up Default section. If the sensors are in
isolation, establish process flow via the Controls menu. See page 48 for the
Moisture valves and page 72 for the Oxygen valves.
7.
Once flow has been established through the moisture cell, observe the pressure
indicator on the main display. For all gases, with the exception of helium, the
pressure in the moisture cell is only controlled by adjusting the needle control valve
on the aspirator. See Table 1 for appropriate pressure settings. In the case of
helium, the vacuum source must be off, and the needle control valve wide open.
8.
It is important to ensure that a minimum of 0.2 slpm flows through the bypass loop
at all times to prevent backflow. See page 31 for additional information on back
pressure prevention.
9.
The proper pressure ranges listed in Table 1 will automatically appear on the
display when the background GSF is selected.
10.
Note: For light gases (hydrogen and helium), the flow indication for the oxygen
sensor will be adjusted automatically by approximately 0.3 to compensate for the
lighter gas. As a result, the flow rate through the oxygen sensor should always be
set to a reading of 1.0 slpm no matter what background gas is in use.
Background Gas
Pressure
Vacuum Source
N2
150 - 250 Torr
ON
Ar
280 - 380 Torr
ON
He
740 - 780 Torr
OFF*
H2
300 - 400 Torr
ON
O2
300 - 400 Torr
ON
*A shut off valve must be installed at the inlet to the aspirator to disable the vacuum.
Table 1: Recommended Sample Outlet Vacuum Pressure
4.2.3 Moisture
Start the gas flow through the Moisture cell by going to the Moisture Controls Menu as
shown on page 47 and select Gas Valves. On the Gas Valves screen Figure 20, highlight
“Restore Process Gas Flow” and hit Apply, and then hit Done.
On initial start-up, after the six minute warm up process is complete, the moisture analyzer
will take a few hours to equilibrate and the concentration of moisture is automatically
shown in parts per million (ppm) or parts per billion (ppb) on the display.
If after initial start-up, the power is briefly turned off, the moisture analyzer will take
approximately six minutes to re-equilibrate.
See the section on Gas Flow and Pressure Regulation on page 29 for important information
on setting proper pressures and flow rates.
Содержание NanoTrace DF-760E
Страница 1: ...NanoTrace Dual Moisture Oxygen Analyzer DF 760E Instruction Manual...
Страница 4: ......
Страница 38: ......
Страница 44: ......
Страница 122: ......
Страница 128: ......
Страница 138: ......
Страница 142: ......
Страница 143: ...Service DF 760E 139...
Страница 144: ...140 DF 760E...
Страница 148: ......
Страница 166: ......
Страница 168: ......
Страница 171: ...DF 760E 167...
Страница 173: ...Appendix A User Menu Screens DF 760E 169 Page 65 Page 66 Page 67 14 2 Oxygen Menus Page 72 Page 74 Page 86 Page 87...
Страница 174: ...170 DF 760E Appendix A User Menu Screens Page 88 Page 91 Page 93 Page 97 Page 98 Page 99...
Страница 175: ...Appendix A User Menu Screens DF 760E 171 14 3 System Menus Page 102 Page 103 Page 103 Page 106 Page 108 Page 109...
Страница 176: ...172 DF 760E Appendix A User Menu Screens Page 110 Page 111 Page 113 Page 114 Page 115 Page 116...