Figure 11 : Calibration-Span menu
The value set in the
Calib Gas Value
column must be the impurity
value of the calibration gas before the dilution. The software
automatically calculates the dilution values based on the dilution
factor.
Calibration enabled
will be displayed in the
RUN-REAL TIME
CHROMATOGRAM
menu.
It is recommended to perform
the
calibration in
Automatic
injection mode. This will ensure better gas
equilibrium.
The range configuration cannot be changed when calibration is
enabled.
Prior to accepting a calibration, ensure the correct gas is flowing
through the analyser and there have been multiple runs to ensure the
readings have stabilised.