Quick Start Guide
SERVO
PRO CHROMA/NANOCHROME/ULTRA Gas Analyser
19
04400003A_0
1.8 Date & Time
Click on Configuration on the menu bar and then on Date & Time to access the
CONFIGURATION
-DATE & TIME menu (see Figure 9).
Use the combo box control to change the value of the date and the time. Click on
Apply
to set the desired value.
Figure 9 : Date & Time menu
1.9 Analyser Calibration
Note:
It is recommended to calibrate the analyser once a month.
Note:
Recommendation is to calibrate with between 80-110% of Full scale range
Click on
Calibration
on the menu bar and then select
Calibration
to access the
CALIBRATION-CALIBRATION
menu (see Figure 10). This menu is used to calibrate the
system for the different gases being analysed.
The analyser must be properly started and purged before the execution of a
calibration. The calibration procedure consists of the Noise Threshold Calibration
followed by the Span Calibration.