10
4.3 Setting Time
1. Speed setting unit: 1 min or 10 min / 1 sec or 10 sec
Time is down-counted after starting centrifugation.
Press the [TIME] button once.
MIN LED is flickering.
2. Press the [▲▼] buttons to change the minute value.
After keeping holding finger on the [▲▼] buttons for
5 seconds, the unit of setting value is changed from
1 min to 10 min. If you do not touch the [▲▼] button
for 5 seconds, the setting mode is cleared.
3. Press the [ENTER] button to pass the sec value
setting.
After keeping holding finger on the [▲▼] buttons for
5 seconds, the unit of setting value is changed from
1 sec to 10 sec. If you do not touch the [▲▼] button
for 5 seconds, the setting mode is cleared.
4. Touch the [Enter] button to complete the setting.
4.4 Setting Temperature and Fast Cool
4.4.1 Setting Temperature
1. Temperature can be set from -20 °C to 40 °C
Temp setting unit: 1 °C
Touch the [TEMP] button.
Temperature value blinks on the display window.
2. Press the [▲▼] buttons to change input value.
If you do not touch the [▲▼] button for 5 seconds,
the setting mode is cleared.
Touch the [Enter] button to save and complete the
setting.