background image

ISOLOK MSA Sampler 13

1. 

Unclamp and remove the sampler from its mounting tee . Remove the gasket .

2. 

Remove any sample container adapter, including the O-ring seal within the body port .

3. 

Disconnect the clamp between the sampler body and barrel . The spring will force the plunger and barrel away 

from the body .

4. 

Pull the plunger and spring out of the barrel . Leave the barrel connected to the air line tubing .

5. 

Remove four plunger seal O-rings from the grooves in the plunger by squeezing the O-ring diameter between 

the thumb and finger and rolling the rings out of their grooves . Two front plunger seal O-rings should be 

rolled forward, off the plunger . Two back seals can be rolled into the sample collecting ring on the plunger for 

convenience during cleaning . Remove the rear rod seal O-ring and piston seal O-ring from the plunger .

6. 

Spray or brush clean/sanitize the body, plunger, spring, all seals and gasket, as well as any port adapter and its seal, 

in approved sanitary procedure .

7. 

Replace the rod seal and piston O-ring into their grooves . Roll the O-ring plunger seal rings into the grooves on 

the plunger . Wipe a light film of approved dairy lubricant onto all six O-ring seals . Add a light film of lubricant on 

the inside of the air cylinder barrel .

8. 

Assemble the spring on the plunger and insert the plunger into the sampler body .

9. 

Slip the barrel over the plunger piston . Press the barrel against the sampler body to compress the spring . Install 

the clamp . Replace any port adapter, including the O-ring seal .

10. 

Replace the line tee gasket and install the sampler and clamp in place . The sampler body should be horizontal, 

with the sample discharge port pointing down . 

Weekly Inspection

With the sampler removed from the line, turn the compressed air and sampler controller power on . Between cycles, 

the plunger dwells retracted . When a cycle is initiated, the sampler plunger should extend 7/8 in (22mm) from the 

sampler body and then retract inside the body . If the seals are completely dry, a sluggish or incomplete stroke may 

occur .

Lubrication

Cleaning normally will remove all lubrication . Wipe a thin film of approved dairy lubricant, such as white grease, CIP 

lube or Vaseline, onto all seals . Continued short stroke, incomplete cycling or sluggish extension may indicate dirt or 

liquid in the barrel, leaks in the air line, or damage to metal parts .

Plunger Seals

Inspect the O-ring seals on the plunger . Replace any that are scuffed, scored or with flats worn on OD . Carefully insert 

the plunger through the bore of the body . Seals should move through smoothly, but with light resistance . Little or no 

resistance indicates the rings are not properly sealing and should be replaced . The bore of the body where the seals 

ride also should be smooth and free of nicks and scratches .

Complete Sampler Disassembly and Maintenance

A complete disassembly, inspection and cleaning of the ISOLOK MSA sampler should be performed as required . Work 

on a clean, well-lighted bench area with a copy of the sampler group assembly drawing for reference .

Содержание ISOLOK MSA

Страница 1: ...Original Instructions Installation Operation Maintenance Manual ISOLOK MSA Sampler Point Samplers S AS IOM 00300 8 10 22...

Страница 2: ...ability Sentry Equipment Corp its employees agents and the authors and contributors to this document specifically disclaim all liabilities and warranties express or implied including warranties of mer...

Страница 3: ...Controller Installation 8 Inspection 11 Operation 11 Initial Cleaning 11 Starting the Sampler 11 Maintenance 12 Daily Cleaning 12 Weekly Inspection 13 Complete Sampler Disassembly and Maintenance 13 C...

Страница 4: ...ons DANGER indicates a hazardous situation which if not avoided will result in death or serious injury WARNING indicates a hazardous situation which if not avoided could result in death or serious inj...

Страница 5: ...at the equipment is isolated from all connecting piping the equipment is depressurized the contents have been drained and the equipment is cool Moving parts This equipment may contain moving parts All...

Страница 6: ...se or piping to pass the sampler before operation begins During initial delay the controller solenoid valve is energized causing the sampler plunger to extend into the mounting tee where it can be flu...

Страница 7: ...y plastic bags that close with a built in wire tie Sizes up to about an 8 oz 250 ml capacity can be supported by carefully folding the open bag mouth and wire around the small diameter below the drip...

Страница 8: ...sampler controller in its most effective location 1 The line where the sampler is to be located can be horizontal vertical or at any angle The location selected must mount the sampler body with its l...

Страница 9: ...ed around a sealing bulkhead wall without the danger of pinching the tube or blocking the door Wall Penetrations The following procedure is suggested whenever air tube must be routed through a wall be...

Страница 10: ...miting device This device automatically disconnects power to the controller in the event of an accidental overload or short circuit Identified as AutoFuse this device opens the power circuit any time...

Страница 11: ...After installation the sampler periodically should be checked for proper operation and to ensure that any wear is detected so proper preventive maintenance can be performed The Maintenance section of...

Страница 12: ...knob with a screwdriver in the slot provided 2 Start delay adjustment 0 1 10 seconds 3 Turn clockwise to increase counterclockwise to decrease delay 4 Start delay is a single turn potentiometer It wi...

Страница 13: ...r body to compress the spring Install the clamp Replace any port adapter including the O ring seal 10 Replace the line tee gasket and install the sampler and clamp in place The sampler body should be...

Страница 14: ...mble the four plunger seals rod seal and piston seal O rings onto the plunger 2 Apply a thin film of the approved dairy lubricant to all six seals Assemble the spring onto the plunger and insert it in...

Страница 15: ...roller sample rate potentiometer is a round component with brown wire leads on two of three lugs If controller operation becomes erratic e g cycles at some load size dial settings but not at others or...

Страница 16: ...alve using the following steps 1 Disconnect the air line tubing to the sampler Disconnect the incoming compressed air supply line Open the sampler controller enclosure Disconnect the wire leads betwee...

Страница 17: ...y monitor and measure processes for improved production efficiency output and safety Standing behind our commitments we are determined to tackle any application anywhere We know that running an effici...

Страница 18: ...r combining both into one connection and using the remaining pin for the sampler power lead AC Power Supply Package This parts package is used with 230 VAC supply and rectifies and transforms 230 VAC...

Страница 19: ...ation NOTE Within the motor junction box either T5 or T8 normally will be connected to T2 and T3 Select a combination of motor leads so that the sampler is NOT wired into this connection Polarity The...

Страница 20: ...d U S Dept of Transportation or equal for service to at least 120 psi 827 KPa Tanks can be obtained locally from firms that build or repair truck air brake systems or an air tank package is available...

Страница 21: ...roduces less than 1 U S ounce 30 cc or 60 sampler cycles A biased sample could result 3 Select a sample container size that will not be filled completely even on longest sampling times Use care in set...

Страница 22: ...inute etc actual delivery will be influenced by many factors some of which are Speed of pump drive and pump condition Type of valve on farm tank Suction hose size and condition Head pressure suction a...

Страница 23: ...ISOLOK MSA Sampler 23 This page is intentionally left blank...

Страница 24: ...sentry equip com 966 Blue Ribbon Circle North Oconomowoc WI 53066 U S A 1 262 567 7256 support sentry equip com Serving customers in more than 50 countries across six continents worldwide...

Отзывы:

Похожие инструкции для ISOLOK MSA