3.3  Cleaning tips 

 

To avoid scratches on the sensor  surface rinse the probe thoroughly using water before cleaning 
the  probe  with the  soft  brush  supplied with the  meter  and  tap  water  with a mild detergent 
added.  
Most scratches on the sensor are caused when there are hard particles in the sample and the 
sample is rubbed into the sensor when cleaning with the toothbrush. 

 

Proteins,  fats  and  oils  may  be  removed  by  scrubbing  in  a  solution  of  Terg-A-Zyme  (Alconox 
company), a pepsin solution or a similar product. Afterwards, rinse thoroughly with deionized 
or distilled water. Cleaning agents are available at Sentron, ask your local dealer. 

  

Do not use hydrofluoric acid, acetone, MEK or similar agents to clean the probe. 

 

Storage 

For short time periods (≤ 2 Days) probes can be stored best “wet” in e.g. a beaker with pH7. Clean the 
probe first with water and possibly a mild detergent. Then place it in a clean container with fresh pH7 
buffer to prevent pollution of the probe directly after cleaning. 

For longer time periods (>2 days) the probe can be stored “dry” in e.g. the shipping box. As well as with 
short term storage, clean the probe first. Then place the protective cap on the probe tip, containing 
one drop of demi-water preventing a dry out of the diaphragm.  

Always revitalize the probe (see section 3.2) before using it again after a long term storage. 

 

 

Содержание ConeFET

Страница 1: ...SI line pH PROBE MANUAL For products ConeFET 3280 010 CupFET 3200 010 LanceFET 2270 010 LanceFET H 2274 010 MicroFET 9270 010 MiniFET 9202 010...

Страница 2: ...he probe as this may cause damage and will void the guarantee All information provided in this manual is applicable at the moment of publication Sentron holds the right to change components functions...

Страница 3: ...of compliance 4 1 2 Guarantee 4 1 3 Specifications 5 2 Use 6 2 1 Installation 6 2 2 Preparing for use 7 2 3 Operating tips 8 3 Cleaning 9 3 1 Daily cleaning 9 3 2 Revitalizing 9 3 3 Cleaning tips 10...

Страница 4: ...n will according to its own judgment either replace or repair a defect probe within 6 months of shipment provided the cause is a defect in material of manufacturing and the defect appeared during corr...

Страница 5: ...5 1 3 Specifications...

Страница 6: ...obe failure or irreversible damage to the probe Samples must be aqueous solutions or semi solids and compatible with the probe s materials If information is required regarding the chemical resistance...

Страница 7: ...page from a new probe is normal and will not affect the lifetime or performance of the probe Subsequently rinse with deionized or distilled water Soak the probe in pH 7 00 buffer or pH 4 00 buffer wit...

Страница 8: ...late based are less susceptible to carbon dioxide contamination than pH 10 00 buffers borax or carbonate based When slope errors occur it usually indicates a failing probe or a contaminated buffer If...

Страница 9: ...e to be cleaned with a brush Place the probe in warm tap water around 60 C 140 F with a mild detergent for 5 minutes stir periodically Scrub the probe tip with a soft brush and water with a mild deter...

Страница 10: ...r Cleaning agents are available at Sentron ask your local dealer Do not use hydrofluoric acid acetone MEK or similar agents to clean the probe 4 Storage For short time periods 2 Days probes can be sto...

Отзывы: