8
Out to “go” (ready to blow) mode from setting menu
*Short cut key: keep pressing together
Running out of batteries and need to replace batteries to new one
*Replace all batteries at a time
Too low temperature to operate
the device. It will apprear below
than -5
℃
(celsius)
DAYS of calibration grace period activated
CA.d and specific numbers left DAYS to calibration lock will be displayed alternatively
Device need to have calibration service before calibration lock
Device is in calibration lock status. Contact technical service provider
COUNTS of calibration grace period activated
CA.t and specific numbers left COUNTS to calibration lock will be displayed alternatively
(001 = 10 COUNTS left)
Device need to have calibration service before calibration lock
Closing to the calibration required DAYS (e.g. d.30 means 30 days left to calibration
required days) It will ONLY apprear warming up status when the device is turned on
Closing to the calibration required COUNTS (e.g. t.99 means 990 test counts left to calibration
required counts) It will ONLY apprear warming up status when the device is turned on
Battery level indication
It will ONLY apprear warming up status when the device is turned on
Device is under calibration grace period (DAYS or COUNTS)
Device need to have calibration service before calibration lock
Flow error: the breath sample was not strong enough
Blow strong and steady during sampling
USB cable is connected to the device
Err1 error: Pressure sensor damaged so contact technical service provider
PC software is connected to the device
Low battery level indication while device is ready to blow status
Status bar is moving and batteries need to be replaced immediately
* Low battery level indication could be displayed under WHITE LEDstatus due
to power consuming on WHITE LED is much bigger than other color of LED
Too high temperatrue to operate
the device. It will apprear above
than 50
℃
(celsius)
Displays on iblow10